ACCESSORIES | ||
|
|
|
K285 Installation with K149
1. Install the K149 to the K148 per the instructions.
2. Loosen the hose clamps of the K285 entirely, place them around the nozzle and tighten them so that the stationary part of the K285 is directly opposite the K149 arm. See Figure C.10
NOTE: The lower hose clamp must be positioned so it does not touch the arm of the K149.
4. Position the moveable section to the desired flux height and tighten the wing nut. Use the center or lower tapped hole depending on the electrical stickout
5. Cut the flux hose to required length and connect as shown.
FIGURE C.10 - K285/K149
K285 installation with the K129
1. Unscrew the hose clamps enough so they can slip over the tip holder clamping nut.
2. Position the stationary portion of the K285 as shown in Figure C.11 and tighten the clamps.
3. Position the moveable section to the desired flux height and tighten the wing nut.
4. Due to the 7° angle of the tips it may be necessary to pivot the cone as shown. Remove the outermost screws from each side of the cone. tilt the cone and put the two screws into the rear of the cone. Tighten all 4 screws.
5. Cut the flux hose to required length and connect as shown.
FIGURE C.11 - K285/K129
IMPORTANT
K285 on ALL Nozzles
After the K285 is installed to any of the compatible nozzles, use an ohmmeter or a test light to insure proper insulation (no continuity) between the copper flux cone and the nozzle body