Understanding the HP/Agilent 8922 GSM HSCSD Test System

Features

Features

The HP/Agilent 8922 GSM HSCSD Test System supports the following features. The full range of features available in speech mode is maintained.

Support of fully variable power level of each active timeslot on the downlink. This provides complete freedom to control the amplitude of the downlinks.

Support of fully independent power control by protocol of each active timeslot on the uplink. This allows you to independently set the Tx level of each timeslot.

Transmit SACCH in each active slot on the downlink (call maintenance + power control) according to ETSI multi-frame structure. This provides a SACCH frame in the correct position on both the downlink TCHs.

All existing measurement functionality and performance maintained and performed on any slot independently. You can select either uplink slot to measure, either by selecting one instrument or via a field on the user interface.

Support of protocol to set up a 2x2 or 2x1, 9.6 or 14.4 kbps timeslot uplink/downlink data channel (both Mobile Originated and Mobile Terminated). The existing HP/Agilent 8922P functionality to originate or receive a call is still available.

Support of burst-by-burst (fast) BER test on both 14.4 kbps timeslot data channels simultaneously. The existing bit error test functionality of the HP/Agilent 8922P is available in speech mode only.

You can select any timeslot for the master channel in the range 2 through 5. The slave timeslot is always the following slot. Any ARFCN can be used in any of the currently supported bands (GSM900, E-GSM, DCS1800 and PCS1900).

Dual-band assignment supported. The existing HP/Agilent 8922P functionality to perform a dual-band assignment is maintained.

All 2x2 HSCSD calls are in transparent mode and all 2x1 calls are in non-transparent mode.

4

Chapter 1

Image 18
Contents HP/Agilent 8922 GSM Hscsd Test System Option #K09 Warranty Limitation of Remedies and Liability Assistance Statement of Compliance Sound Emission Manufacturer’s DeclarationEarthing ground, incorporated in the power cord. The mains Instrument dangerous. Intentional interruption isProhibited Safety Symbols Safety SymbolsISM 1-A Overview User’s Guide AbbreviationsContents Remote operation Returning equipment for repair Upgrading the HOP Controller ROM and Firmware Understanding the HP/Agilent Getting an overview of the HP/Agilent GSM Hscsd Test System HP/Agilent 8922 GSM Hscsd Test System, option #K09Chapter Features Setting up the HP/Agilent Before you start Checking the HP/Agilent 8922 Requirements for Hscsd Making the front panel connections Front panel connectionsMaking the rear panel connections StepRear panel connections Enabling the Hscsd functionality Inserting the keycard into the HP/AgilentTests screen Config screen Calibrating the overall system Manual operation Configuring the system Message Frame reset sent appears in the top left-hand corner Establishing a call Making measurements Choosing which timeslot to measureMeasuring BER Changing power levels, timeslot and channels Hscsd control screenEnding a call Ending a call Screens Config screen Instrument typeChapter Slave address Hscsd screen Master downlink amplitudeSlave offset Master channelSelect ‘Master’ to measure before signalling mobile Master timeslot Slave timeslotCall Type Data rate MeasureMaster Tx level Master input level Master mode Slave Tx levelSlave input level Slave control Peak power display Mobile reportsTo Screen menu Fast BER Screen Fast BER screenLoopback Remote operation Introduction Configure Subsystem Gpib Commands CONFigureINSTrumentTYPEConfigure Instrument Type comand syntax Display Subsystem Gpib Commands Display Screen command syntaxDISPlaySCReen Hscsd Hscsd Subsystem Gpib Commands Hscsd command syntax master unit Hscsd command syntax slave unit SYNC? TCHHSCSdMASTerTCHARFCn HSCSdMASTerSTATusTCHTSLot Command 0ptions HSCSdMASTerDOWNlinkAMPLitudeHSCSdMASTerTCHTLEVel HSCSdMASTerUPLinkAMPLitude HSCSdMASTerUPLinkAMPLitudeCONTrol HSCSDMASTerCALLTYPE HSCSDMASTerCALLRATE HSCSdSLAVeSTATusSYNC HSCSdSLAVeSTATusTCHTSLot? HSCSdSLAVeDOWNlinkAMPLitudeOFFSet HSCSdSLAVeTCHTLEVel HSCSdSLAVeUPLinkAMPLitude HSCSdSLAVeUPLinkAMPLitudeCONTrol Remote operating procedures Configuring the systemEstablishing a call Making measurementsChanging channels, timeslot and power levels Ending a call Example code showing a dual-band handover from E-GSM to DCS Clear ScreenPause Example code showing a dual-band handover from E-GSM to DCS Returning equipment for repair Return Procedure Chapter Sales and Service Offices Canada Sales and Service Offices Upgrading the HOP Controller ROM and Firmware Upgrading the Hop Controller ROM ROMSide View Figure A-3 Removing the A33 Hop Controller Board Figure A-4 Removing the A33 U62 HOP Controller ROMUpgrading the Hop Controller ROM Upgrading the firmware