Setting up the HP/Agilent 8922 GSM HSCSD Test System
Checking the HP/Agilent 8922 Requirements for HSCSD
Checking the HP/Agilent 8922 Requirements for HSCSD
To run the HSCSD functionality you need two HP/Agilent 8922 P systems (one acting as a master and one acting as a slave). Each system must be running firmware version C.01.06 or later and contain a HOP Controller ROM with a revision number of 19980730 or greater.
To check which version of firmware is running on the HP/Agilent 8922, go to the Config screen.
To check which version of HOP Controller ROM is installed in the HP/ Agilent 8922, go to the Service screen and select the hop_revision latch.
For details on how to upgrade the HP/Agilent 8922’s HOP Controller ROM and firmware, see Appendix A , “Upgrading the HOP Controller ROM and Firmware,” on page 75.
Chapter 2 | 7 |