Manual operation
Configuring the system
| Configuring the system |
| To configure the HP/Agilent 8922 GSM HSCSD Test System for manual |
| operation, proceed as follows. |
| You must configure the slave unit first as a frame reset is sent from the |
NOTE | |
| master unit to the slave unit as soon as the master unit is activated. The |
| slave unit must be ready to receive this frame reset. |
|
|
NOTE | If you are operating the HP/Agilent 8922 GSM HSCSD Test System |
| manually do not connect another GPIB controller such as a PC to the |
| GPIB bus. The master unit operates as the controller and any other |
| device attached to the GPIB bus can result in conflicts on the GPIB bus. |
|
|
Step 1. On the slave unit, navigate to the CONFIG screen and set the Instrument Type to HSCSD Slave.
Figure | Config screen |
| |
|
|
|
|
|
|
|
|
|
|
|
|
C.01.06
16 | Chapter 3 |