PERFORMANCE
Reference Conditions
Generally as
Ambient Temperature: | 20C 2oC | |
Relative Humidity: | 60 - | 70% |
Supply Voltage: | 90 - | 264V AC 50Hz 1% |
Source Resistance: | <10 for thermocouple input | |
Lead Resistance: | <0.1/lead balanced (Pt100) |
Performance Under Reference Conditions
Common Mode Rejection: | >120dB at 50/60Hz giving negligible |
| effect at up to 264V 50/60Hz. |
Series Mode Rejection: | >500% of span (at 50/60Hz) causes |
| negligible effect. |
DC Linear Inputs |
|
Measurement Accuracy: | 0.25% of span 1LSD. |
Thermocouple Inputs |
|
Measurement Accuracy: | 0.25% of span 1LSD. NOTE: Reduced |
| performance with Type “B” Thermocouple |
| between 100 - 600C (212 - 1112F). |
Linearisation Accuracy:
Cold Junction
Compensation:
RTD Inputs
Better than 0.2C any point, any 0.1C range (0.05C typical). Better than 0.5C any point, any 1C range.
Better than 0.7C.
Measurement Accuracy:
Linearisation Accuracy:
0.25% of span 1LSD
Better than 0.2C any point, any 0.1C range (0.05C typical). Better than 0.5C any point, any 1C range.
Volume II |