5.DMM #1 should read 4.995 V, ±.01 V

6.If the reading of step 5 is out of tolerance, adjust potentiometer R35 on CPU Board to correct.

7.Recheck both points and readjust if required.

5.8.2 Current Output Span & Load

The current output circuitry is provided with two calibration adjustments. Potentiometer R50 is used for span adjustment and R49 for load compensation. For these tests DMM #2 using the current measurement function is connected in series with a 250 ohm resistor load as shown in Figure 5-3. The pushbutton switch is used to jump the load.

1.Adjust input test circuit for 4.995 V on DMM #1.

2.DMM #2 should read 20 mA. If necessary, adjust pot R50 to correct.

3.Press load pushbutton while observing DMM #2. If current reading increases, adjust R49 CCW. If reading decreases, turn it CW. Repeat procedure until change is less than

±.02 mA.

4.Final reading should be 20 mA, ±.02 mA. If necessary, reset pot R50 and repeat procedure.

5.Calibration is complete. Restore unit to normal operating status.

5.9 PCB REFERENCE DRAWINGS

Table 5-1 below provides part numbers and drawing references (Figures 5-4 through 5-7) for the printed circuit boards provided with the Remote Set Regulator.

Table 5-1 - CPU Board & Termination Board Reference Table

Fig.

Assembly

Part #

Pag

 

 

 

e

5-4

Assembly, Analog Type CPU Board w/ Current Output (12V)

389615-03-0

5-9

5-4

Assembly, Analog Type CPU Board w/ Current Output (24V)

389615-06-4

5-9

5-5

Assembly, Analog Termination Board

389644-01-3

5-9

5-6

Assembly, Raise/Lower Type CPU Board w/ Current Output (12V)

389613-03-7

5-10

5-6

Assembly, Raise/Lower Type CPU Board w/ Current Output (24V)

389613-06-1

5-10

5-7

Assembly, Raise/Lower Termination Board

389616-01-0

5-10

Service

5-8

CI-9110

Image 48
Contents Remote Set Regulator Model 9110-00A IMPORTANT! Read Instructions Before Starting Warranty Mail Bristol IncPhone Bristol Inc. Repair Authorization Form off-line completion Training Telephone Support Technical Questions 860 860 945-2213 FAXVisit our Site on the World Wide Web Training Courses Model 9110-00A Remote SET Regulator Supplemental Instructions Product Description Theory of OperationAnalog Regulator Model Remote Set Regulator Block Diagram of Analog Type Regulator Guard Input Current OutputBlock Diagram of Raise/Lower Type Regulator Data Plate Raise/Lower Regulator ModelModel Number Breakdown = Pneumatic Output = Mounting Arrangement Page General MountingTwo-Inch Pipe Mounting In-Line MountingPressure Connections General PipingRegulator Assembly Moisture in LinesInternal Access Line SizeVenting Body and CoverSafety Note Electrical Conduit PortPage Using Option Switches Cradle Switches with Open DesignationSlide Switches with on Designation Switch Functions Regulator Assembly Switch SW1 & SW2 Functions & SettingsAnalog Input Models RAISE/LOWER Input ModelsContinuous Signal Incremental SignalSwitch SW1 1-5, Pulse Increments Input FilterGuard Input Output Rate of ChangeStatic or Dynamic Guard Input Guard LogicSwitch SW2 2-5, Output Rate Of Change Analog Failure Modes Fail Hold & Fail ZeroFAIL-ZERO Mode Page Field Wiring Terminal Identification For Analog Input Models Terminal Identification For Raise/Lower ModelsAnalog Voltage Input with Guard Analog Current Input 4-20 mA with Guard RAISE/LOWER Inputs with GuardRaise/Lower & Guard Inputs from Relay Contacts DC Power Current Output to FSK Bi-State TransmitterOperating Notes Initial ChecksField Wiring CI-9110 Service Troubleshooting Hints Setting Output Limits Test SetupProcedure General PrinciplesCalibration Test Setup Calibration PrecautionAnalog Test Setup Preliminary Calibration CheckDetailed Calibration for Analog Model Feedback Zero & SpanCurrent Output Span & Load Pressure Output CheckRaise/Lower Test Setup Detailed Calibration for RAISE/LOWER ModelPCB Reference Drawings CPU Board & Termination Board Reference TableAssembly, Analog Type CPU Board with Current Output Assembly, Raise/Lower Type CPU Board with Current Output Analog Input Model RAISE/LOWER Input ModelAnalog Output Accuracy Analog ModelsAccuracy RAISE/LOWER Models Stepper Drive MechanismPressure Regulator Assembly Instrument EnclosureEnvironmental Range Factory Calibrated withHumidity 10 to 95% over -20to +130 F -29to +55 Range Model 9110 Main Parts Parts Referenced in FigureRemote Set Regulator Actuator Unit Remote Set Regulator Actuator Unit Mounting Parts Remote Set Regulator Assembly Basic Parts Remote SET Regulator Basic PartsGuidelines for System Grounding Basic Ground Rod Installation Noise and Signal Errors General RecommendationsGrounding of Equipment Housing Transients and Interference Grounding of Transmitter Types of Earth Grounds AC Power Grounding System Dry Soil Sandy SoilRocky Soil Several DPCs Mounted in Metal Cabinet with Power Supply Following installation procedures applyGrounding for DPCs and Power Supplies Single Metal Cabinet Multiple DPC Cabinets Powered by Single Power Supply Elements of a Facility Ground System Multiple Clusters of DPC Cabinets Powered by Local Supplies ReferencesPage Remote Set Regulator Model 9110-00A