Hysteresis:

 

(Output to gauge) 0.5% of span

Motor Protection:

 

Adjustable mechanism limits restrict output

 

 

for over and under range conditions.

 

 

Automatic timeout prevents clutch wear.

Power Failure:

 

Mechanism maintains output pressure prior

 

 

to loss or drop of supply voltage.

6.9 PRESSURE REGULATOR ASSEMBLY

Output Ranges Offered:

 

Range

Factory Calibrated with

 

 

3 - 15 psig

 

20 psig supply

 

3 - 27 psig

 

30 psig supply

 

6 - 30 psig

 

35 psig supply

Supply Pressure Range:

 

Type

Min.

Max.

 

Unit

Supply

Supply

 

(psig)

(psig)

(psig)

 

3 - 15

20

100

 

3 - 27

30

100

 

6 - 30

35

100

Supply Pressure Effect:

 

0.1 psig Max. change in output for each 10

 

 

psig change of input.

Connection:

IN, OUT and VENT ports (1/4 Inch NPT)

6.10 INSTRUMENT ENCLOSURE

 

 

Type of Case:

 

Weatherproof and explosion proof. Cast

 

 

aluminum with gasketed screw-on cover.

 

 

Gray epoxy finish. Meets NEMA-3

 

 

specifications

 

General Dimensions:

6 inch (152.4 mm) diameter

 

 

7 inch (176.8 mm) length

Weight:

 

10 pounds (4.54 kg)

 

Mounting Options:

 

In-line pressure pipe or 2 inch pipe bracket

6.11 ENVIRONMENTAL

 

 

 

 

Temperature Limits:

 

-20°to +150° F (-29°to +65° C) Operating

 

 

-40°to +185° F (-40°to +85° C) Storage

CI-9110

6-3

Specifications

Image 53
Contents Remote Set Regulator Model 9110-00A IMPORTANT! Read Instructions Before Starting Warranty Phone MailBristol Inc Bristol Inc. Repair Authorization Form off-line completion Training 860 860 945-2213 FAX Telephone Support Technical QuestionsVisit our Site on the World Wide Web Training Courses Model 9110-00A Remote SET Regulator Supplemental Instructions Analog Regulator Model Product DescriptionTheory of Operation Remote Set Regulator Block Diagram of Analog Type Regulator Current Output Guard InputBlock Diagram of Raise/Lower Type Regulator Model Number Breakdown Data PlateRaise/Lower Regulator Model = Pneumatic Output = Mounting Arrangement Page Mounting GeneralTwo-Inch Pipe Mounting In-Line MountingGeneral Piping Pressure ConnectionsMoisture in Lines Regulator AssemblyLine Size Internal AccessVenting Body and CoverElectrical Conduit Port Safety NotePage Slide Switches with on Designation Using Option SwitchesCradle Switches with Open Designation Regulator Assembly Switch SW1 & SW2 Functions & Settings Switch FunctionsAnalog Input Models RAISE/LOWER Input ModelsIncremental Signal Continuous SignalInput Filter Switch SW1 1-5, Pulse IncrementsOutput Rate of Change Guard InputStatic or Dynamic Guard Input Guard LogicAnalog Failure Modes Fail Hold & Fail Zero Switch SW2 2-5, Output Rate Of ChangeFAIL-ZERO Mode Page Field Wiring Terminal Identification For Raise/Lower Models Terminal Identification For Analog Input ModelsAnalog Voltage Input with Guard RAISE/LOWER Inputs with Guard Analog Current Input 4-20 mA with GuardRaise/Lower & Guard Inputs from Relay Contacts Current Output to FSK Bi-State Transmitter DC PowerInitial Checks Operating NotesField Wiring CI-9110 Service Troubleshooting Hints Test Setup Setting Output LimitsProcedure General PrinciplesCalibration Precaution Calibration Test SetupPreliminary Calibration Check Analog Test SetupFeedback Zero & Span Detailed Calibration for Analog ModelCurrent Output Span & Load Pressure Output CheckDetailed Calibration for RAISE/LOWER Model Raise/Lower Test SetupCPU Board & Termination Board Reference Table PCB Reference DrawingsAssembly, Analog Type CPU Board with Current Output Assembly, Raise/Lower Type CPU Board with Current Output RAISE/LOWER Input Model Analog Input ModelAccuracy Analog Models Analog OutputAccuracy RAISE/LOWER Models Stepper Drive MechanismInstrument Enclosure Pressure Regulator AssemblyEnvironmental Range Factory Calibrated withHumidity 10 to 95% over -20to +130 F -29to +55 Range Parts Referenced in Figure Model 9110 Main PartsRemote Set Regulator Actuator Unit Remote Set Regulator Actuator Unit Mounting Parts Remote SET Regulator Basic Parts Remote Set Regulator Assembly Basic PartsGuidelines for System Grounding Basic Ground Rod Installation General Recommendations Noise and Signal ErrorsGrounding of Equipment Housing Transients and Interference Grounding of Transmitter Types of Earth Grounds AC Power Grounding System Rocky Soil Dry SoilSandy Soil Following installation procedures apply Several DPCs Mounted in Metal Cabinet with Power SupplyGrounding for DPCs and Power Supplies Single Metal Cabinet Multiple DPC Cabinets Powered by Single Power Supply Elements of a Facility Ground System References Multiple Clusters of DPC Cabinets Powered by Local SuppliesPage Remote Set Regulator Model 9110-00A