Loading the machine
Depending on the type of labware you wish to clean, the washer can be used with an upper basket, lower basket, or injector wagon, as well as an assortment of inserts and special baskets to suit your needs.
Baskets and inserts should be properly selected to match the application desired.
After every program completion, conduct the following visual inspection:
–Inspect the labware for cleanliness.
–Has any hollowware been dislodged from its proper position on the nozzles?
,Any items that were dislodged from the adapters during a program cycle must be washed again.
Before every program start, conduct the following visual inspection:
–Is the labware properly loaded and secured in the machine?
–Are the hollow vessels properly loaded so that all cavities and channels will be flushed clean?
–Are the spray arms clean, and can they rotate freely?
–Is the filter combination free of coarse debris (remove any coarse material and clean the coarse, fine, flat, and
–Is the basket adapter properly connected to the water supply for the spray arms and nozzles?
–Are the chemical dispensers filled?
–Are the cavities and channels in any hollow vessels cleaned through?
–Are the nozzles and connections firmly attached to the baskets/inserts?
24