Plumbing
Connecting the water inlet
,The water in the machine is not drinkable.
–The machine must be connected to the water supply in accordance with local and national regulations.
–The washer is constructed so it may be connected to a suitable supply without an extra
–High iron content can leave a rust film on stainless steel labware and the machine itself.
When the chloride content in the domestic water supply is above 100mg/l, the risk of corrosion damage to stainless steel labware is greatly increased.
–In certain regions (e.g., mountainous areas) the water composition may cause condensate to form, requiring the use of softened water in the steam condenser.
–An acceptable water pressure (flow rate) is 10 - 147 psi. However, the recommended pressure is
25 - 60 psi.
If the water pressure is below 30 psi, the fill time will take longer.
If the "Fill/Drain" indicator fault code "F..E" lights, contact the Miele Technical Service Department.
–The maximum permissible static water pressure is 145 psi over pressure.
–If the water pressure lies outside these ranges, please contact Miele Technical Service for advice (the "Fill/Drain" indicator, may appear in the display "F E").
–The machine is designed for connection to cold water (blue marking) and hot water (red marking) up to 158°F (70°C) max.
Connect the water intake hoses to the cold and hot water faucets.
–If there is no hot water supply, the intake hose with the red marking must also be connected to the cold water supply.
69