Miele G 7883 CD DOS 1 liquid detergent, Programming functions, Setting the detergent dosage, Step

Page 37
DOS 1 (liquid detergent)

Programming functions

DOS 1 (liquid detergent)

Setting the detergent dosage

The dosage concentration only needs to be set for liquid detergents. Set the concentration according to the detergent manufacturer's recommendations. Do not change the setting if you are using powdered detergent in the door dispenser.

Step

Display / Result

^Press the 0 button until E02 appears in E02 the display.

^Turn the program selector to "Inorganica".

Using door dispenser

10 = 10 sec dosage length

Using DOS module

1.00 = 1.0 % dosage concentration

^ Press or hold down the ß button until

0.50 The dosage time / dosage

the required value appears in the

concentration is now set.

display, e.g., 0.5.

 

 

 

 

 

 

Dosage concentration setting:

 

 

 

 

 

 

Priming the dosage system

 

 

The dosage system does not need to be primed (cleared of air) unless:

the dosage concentration is being set for the first time,

the liquid detergent container was inadvertently completely emptied .

Step

Display / Result

^ Press the 0 button until E02 appears

E02

in the display.

 

^Turn the program selector to "Custom". d§1

^ Press the ß button.

The dosage system will be primed

 

automatically. Priming is complete

 

when 0 appears in the display.

^Exit the programming mode. Start "Rinse" program.

This will rinse away any agents that entered the wash cabinet during priming.

37

Image 37
Contents Laboratory Glassware Washer G 7883 CD Operating and Installation Instructionsen - US, CA Page Contents Program Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Programming functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 The following instructions apply to all programming functions:. . . . . . . . . . . . . . . 36 To access the programming mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Saving to memory and exiting the programming mode . . . . . . . . . . . . . . . . . . 36 Exiting the programming mode without saving . . . . . . . . . . . . . . . . . . . . . . . . 36 DOS 1 liquid detergent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Setting the detergent dosage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Priming the dosage system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 DOS 3 neutralizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Setting the neutralizer dosage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Priming the dosage system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Changing program parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Changing the temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Changing the holding time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Restoring the factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Setting the time and date. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Resetting the coarse and fine filter timer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Data transfer optional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Routine checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Cleaning and Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Frequently asked questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Thermo switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 After Sales Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 INSTALLATION INSTRUCTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Plumbing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 Caring for the environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IMPORTANT SAFETY INSTRUCTIONS ~Take care when handling processing chemicals detergents, softeners, neutralizers, etc.. These may contain irritant or corrosive ingredients. Do not use organic solvents as these could cause an explosion. Wear protective gloves and goggles. With all chemical agents the manufacturer’s safety conditions must be observed IMPORTANT SAFETY INSTRUCTIONS Accessories Please note the following symbols on the machineSAVE THESE INSTRUCTIONS Disposal of an old applianceGuide to the lab washer a Drying unit h Service panel i Serial interfacec Dispenser for powdered detergent d Rinse aid reservoir with dosage selectorStandard model Description of the machineIntended applications: Laboratory equipment Electric door lock Opening and closing the doorTo open the door To close the doorWater softener Setting the water softenerfactory setting SettingsFilling the salt reservoir Salt indicator Adding rinse aid Rinse AidSetting the rinse aid dosage Rinse aid indicator lightAdding neutralizer Neutralizing agentNeutralizer indicator light Detergent Setting the liquid detergent dosageAdding detergent Liquid detergent refill light ^Snap the cover closed ^Load the detergent into the compartmentLoading the machine Prepare items before loading Loading the machineSpecial preparations before operating the machine Spring adapter for the water connection Adjusting the upper basketTo adjust the upper basket Laboratory equipment For various inserts Upper basket OLower basket U O Injector wagon E Injector wagonFor narrow necked glassware Upper basket For direct injection ofStarting a program Turning onSelecting the optional drying function Program sequence display Turning off End of a programAdditional drying Canceling a program Program Guide Program sequence To access the programming mode Programming functionsGeneral information Saving to memory and exiting the programming modeSetting the detergent dosage DOS 1 liquid detergentStep Display / ResultSetting the neutralizer dosage DOS 3 neutralizerBuzzer Changing the wash and rinse parameters Changing program parametersChanging the drying parameters Changing the holding time Step Changing the temperature StepRestoring the factory settings Setting the time and date Resetting the coarse and fine filter timer Data transfer optional Routine checks MaintenanceCleaning the filters in the wash cabinet Cleaning and CareCleaning the fine, flat and micro-finefilters Removing the spray arms Cleaning the spray armsCleaning the drain pump and non-returnvalve Cleaning the filters Cleaning the water inlet filtersCleaning the exterior Cleaning the control panelCleaning the wash cabinet Cleaning the door sealBaskets and inserts Possible fault Frequently asked questionsSolution What ifWhat if Frequently asked questionsWhat if Frequently asked questionsWhat if Frequently asked questionsWhat if Frequently asked questionsWhat if show signs of Items made of glasscorrosion Items made of stainlessdeposits and/or spots There are whiteon the labware unacceptable The wash result isThermo switch V1-800-565-6435service@miele.ca After Sales ServiceU1-800-991-9380techserv@mieleusa.com INSTALLATION INSTRUCTIONS Leveling the legs InstallationPositioning and securing the machine Protecting the countertop Removing the lid if necessaryCover plate protects the countertop THIS APPLIANCE MUST BE GROUNDED Electrical connectionGrounding connection Plumbing Connecting the water inletPlease also refer to the installation diagram DI connection unpressurized optional DI Connection pressurizedDrainage English Technical dataMetric Disposal of the packing materials Caring for the environmentDisposal of an old appliance Page M.-Nr.07 157 440 / Subject to change without notice /