Miele G 7883 CD installation instructions Important Safety Instructions

Page 7
Manual background

IMPORTANT SAFETY INSTRUCTIONS

The following points should be observed to assist in maintaining quality standards for critical labware and to avoid damage to the loads being cleaned.

~For this type of cleaning, the program must not be interrupted until completion. In other words, the On/Off Button must not be used. Interruptions should also be avoided for all other programs, such as during performance testing, because this can negatively affect the cleaning and rinsing results. After any interruption, the program should be restarted from the beginning.

~Only use cleaning agents formulated for special processes and approved by Miele for use with this machine. Use of unsuitable cleaning agents could adversely affect the components of the machine.

~Use only Miele approved cleaning agents with this machine. Use of unsuitable cleaning agents could adversely affect the components of the machine. Damages resulting from using unsuitable cleaning agents are not covered by the warranty.

~Pre-treatments with cleaning agents can create suds, as can certain rinsing agents. For pre-treatment and / or cleaning, only use low-sudsing detergents which have been approved by Miele. Suds can have an adverse effect on the cleaning process.

~If toxic chemicals may be present in the wash cabinet, beware of this risk when interrupting a program and opening the washer. The door seal and steam condensor must always be checked for soundness.

~The machine must only be operated with water and the recommended cleaning agents. Organic solvents must not be used in the machine as there is the danger, under certain circumstances, of explosion and damage to rubber and synthetic materials. Follow the dosage recommendations of the cleaning agent’s manufacturer.

~In critical applications where very stringent requirements have to be met, it is strongly recommended that all the relevant factors for the process, such as detergent, quality of water, etc. are discussed with a Miele Application Technology specialist.

~It is the responsibility of the operator to routinely check that the required cleaning standards are being met. Therefore, results need to be regularly tested and documented, both thermo electrically and through inspection. For thermo chemical processes, additional testing is required using chemical or biological indicators.

~The performance of the cycles was tested using Neodisher products. The use of other products is not discouraged but may not yield identical results. Please be aware that changes in formulation, storage conditions, etc. which may not be publicized by the chemical manufacturer can have a negative effect on cleaning results. and are not the responsibility of the machine manufacturer.

7

Image 7
Contents Laboratory Glassware Washer G 7883 CD Operating and Installation Instructionsen - US, CA Page Contents Program Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Programming functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 The following instructions apply to all programming functions:. . . . . . . . . . . . . . . 36 To access the programming mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Saving to memory and exiting the programming mode . . . . . . . . . . . . . . . . . . 36 Exiting the programming mode without saving . . . . . . . . . . . . . . . . . . . . . . . . 36 DOS 1 liquid detergent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Setting the detergent dosage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Priming the dosage system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 DOS 3 neutralizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Setting the neutralizer dosage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Priming the dosage system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Changing program parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Changing the temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Changing the holding time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Restoring the factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Setting the time and date. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Resetting the coarse and fine filter timer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Data transfer optional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Routine checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Cleaning and Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Frequently asked questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Thermo switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 After Sales Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 INSTALLATION INSTRUCTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Plumbing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 Caring for the environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IMPORTANT SAFETY INSTRUCTIONS ~Take care when handling processing chemicals detergents, softeners, neutralizers, etc.. These may contain irritant or corrosive ingredients. Do not use organic solvents as these could cause an explosion. Wear protective gloves and goggles. With all chemical agents the manufacturer’s safety conditions must be observed IMPORTANT SAFETY INSTRUCTIONS Disposal of an old appliance Please note the following symbols on the machineAccessories SAVE THESE INSTRUCTIONSGuide to the lab washer d Rinse aid reservoir with dosage selector h Service panel i Serial interfacea Drying unit c Dispenser for powdered detergentStandard model Description of the machineIntended applications: Laboratory equipment To close the door Opening and closing the doorElectric door lock To open the doorWater softener Setting the water softenerfactory setting SettingsFilling the salt reservoir Salt indicator Adding rinse aid Rinse AidSetting the rinse aid dosage Rinse aid indicator lightAdding neutralizer Neutralizing agentNeutralizer indicator light Detergent Setting the liquid detergent dosageAdding detergent Liquid detergent refill light ^Snap the cover closed ^Load the detergent into the compartmentLoading the machine Prepare items before loading Loading the machineSpecial preparations before operating the machine Spring adapter for the water connection Adjusting the upper basketTo adjust the upper basket Laboratory equipment For various inserts Upper basket OLower basket U Upper basket For direct injection of E Injector wagonO Injector wagon For narrow necked glasswareStarting a program Turning onSelecting the optional drying function Program sequence display Turning off End of a programAdditional drying Canceling a program Program Guide Program sequence Saving to memory and exiting the programming mode Programming functionsTo access the programming mode General informationDisplay / Result DOS 1 liquid detergentSetting the detergent dosage StepSetting the neutralizer dosage DOS 3 neutralizerBuzzer Changing the wash and rinse parameters Changing program parametersChanging the drying parameters Changing the holding time Step Changing the temperature StepRestoring the factory settings Setting the time and date Resetting the coarse and fine filter timer Data transfer optional Routine checks MaintenanceCleaning the filters in the wash cabinet Cleaning and CareCleaning the fine, flat and micro-finefilters Removing the spray arms Cleaning the spray armsCleaning the drain pump and non-returnvalve Cleaning the filters Cleaning the water inlet filtersCleaning the door seal Cleaning the control panelCleaning the exterior Cleaning the wash cabinetBaskets and inserts What if Frequently asked questionsPossible fault SolutionWhat if Frequently asked questionsWhat if Frequently asked questionsWhat if Frequently asked questionsWhat if Frequently asked questionsWhat if Items made of stainless Items made of glassshow signs of corrosiondeposits and/or spots There are whiteon the labware unacceptable The wash result isThermo switch V1-800-565-6435service@miele.ca After Sales ServiceU1-800-991-9380techserv@mieleusa.com INSTALLATION INSTRUCTIONS Leveling the legs InstallationPositioning and securing the machine Protecting the countertop Removing the lid if necessaryCover plate protects the countertop THIS APPLIANCE MUST BE GROUNDED Electrical connectionGrounding connection Plumbing Connecting the water inletPlease also refer to the installation diagram DI connection unpressurized optional DI Connection pressurizedDrainage English Technical dataMetric Disposal of the packing materials Caring for the environmentDisposal of an old appliance Page M.-Nr.07 157 440 / Subject to change without notice /