Teledyne 2230R instruction manual Process Connection

Page 10

MODEL 2230R, Process Hydrogen Analyzer

3.3PROCESS CONNECTION

Model 2230R offer a variety of fittings to mate the unit to a process stream. The following table lists our standard fitting selections. Others are available upon request.

½” MNPT thread

½” FNPT thread

-8 SAE/MS thread size

... and many other industry standards.

3.4PROCESS CONNECTION

Power (8 to 13 VDC) is connected via the Power/Analog connector as shown in Section 4.1. Once power is applied, the unit executes a warm-up sequence lasting five minutes. The status LED will be amber in color during the warm-up sequence. When the unit is ready for operation, the status LED will change to green (if the measured hydrogen concentration is under the first relay set point) and to red (if it exceeds the second relay set point). The following operations will be completed in this warm-up sequence:

Heat the sensor to operating temperature

Perform system self-test

WARNING: MODEL 2230R PROCESS HYDROGEN ANALYZERS ARE CALIBRATED IN A HYDROGEN/NITROGEN (H₂/N₂) BACKGROUND AND SHOULD NEVER BE OPERATED IN AN AIR OR OXYGEN BACKGROUND.

After warm-up, the Model 2230R Analyzer should be run in a hydrogen concentration exceeding 5% by volume for at least an hour. Oxygen will readily adsorb on the Model 2230R Analyzer and can confound the hydrogen measurements. If the sensors are left in oxygen, air or any environment without hydrogen for long periods of time, they must be conditioned in hydrogen to remove adsorbed oxygen and taken through the Verification process to check accuracy. Failure of the Verification will require that the sensors go through Calibration. Both procedures are described later in this manual.

The Model 2230R Analyzer can be exposed to oxygen for short periods of time without adverse effects if the unit is turned off. If the units are operated in oxygen or stored in air for longer than a week, the units’ hydrogen readings may be high due to oxygen adsorption on the sensor. Hydrogen exposure will then cause the readings to drift lower as adsorbed oxygen is slowly removed and the unit recovers to normal steady behavior.

If this drift behavior is observed, the sensor should be conditioned by operating in a hydrogen concentration exceeding 5% until the readings are stable. The required hydrogen conditioning may vary from several hours to several days depending on the level of oxygen exposure. Higher hydrogen concentrations used during conditioning may accelerate the process. Once stable, the unit should be taken through Verification to check accuracy and Calibration if needed.

Page 10

Image 10
Contents Important Notices 30/2011 ECO #Copyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Process Hydrogen Analyzer Model 2230R DescriptionTeledyne Analytical Instruments Teledyne InstrumentsSpecifications Dimensions WeightModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Hydrogen Detected above R2 set point/ Unit fault detected Status Indicator ColorNormal operation / Hydrogen Detected below R1 set point Electrical & Communication Connections Power/Analog Output ConnectorWire Color DescriptionAnalog Output Power-On Self Diagnostic Error Voltage Analog Output RangeExample Display₂ Messages 0.0000 0.0000 Example ESC Format fmt Fmt opt Low high Low high err not rdy AtmFormat Bit Value Identifiers Options Bit Value Identifiers Calculation Errors Heater StateOptions opt Bit value HexadecimalTwo place hexadecimal value Fmt Two Place Hexadecimal Value Command prompt G fmt opt = g 8bOpt Two Place Hexadecimal Value Field Maintenance CleaningGas Connection Accuracy Drift/week Tolerance Chart Gas Temperature 65cVerification Verification IntervalVerification Procedure If the unit fails Verification, you will see the following StepVerification complete An average will be calculatedCalibration Calibration ProcedureCalibration Interval Field Calibration Steps Model 2230R, Process Hydrogen Analyzer