MODEL 2230R, Process Hydrogen Analyzer

3. OPERATION

3.1UNIT LOCATION

The unit can be mounted in any orientation or position, however vertical mounting (unit above process connection) should be made in process streams containing liquids or potentially condensing gases.

3.1.1 WARNING LABEL

Remove sensor tube warning label prior to using the unit.

3.2MOUNTING

Mounting is achieved by securing the sensor tube into the supplied fitting directly in the process piping as shown below. Optional mounting brackets that attach to the instrument housing are also available.

WARNING: DO NOT CINCH DOWN OR TIGHTEN FERRULES OUTSIDE OF THE FERRULE REGION OF THE LONG TUBE OR YOU WILL RISK PERMANENTLY DAMAGING THE LONG TUBE AND SENSOR ASSEMBLY WITHIN.

1.0inches is the maximum distance that the fitting mount can be from the end plate of the Model 2230R unit. This is referred to as the Ferrule Region of the long tube which has a wall thickness of 0.065 in. Any distance exceeding 1.0 in. will be in the Sensor Assembly Region where the tube wall thickness is only 0.038 in. Cinching down a fitting outside the Ferrule Region and in the Sensor Assembly Region may result in permanent damage to the long tube and the sensor assembly within.

Page 9

Image 9
Contents 30/2011 ECO # Important NoticesCopyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Teledyne Analytical Instruments DescriptionProcess Hydrogen Analyzer Model 2230R Teledyne InstrumentsDimensions Weight SpecificationsModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Status Indicator Color Hydrogen Detected above R2 set point/ Unit fault detectedNormal operation / Hydrogen Detected below R1 set point Wire Color Power/Analog Output ConnectorElectrical & Communication Connections DescriptionAnalog Output Voltage Analog Output Range Power-On Self Diagnostic ErrorDisplay Example₂ Messages 0.0000 0.0000 Example ESC Fmt opt Low high Low high err not rdy Atm Format fmtFormat Bit Value Identifiers Options opt Calculation Errors Heater StateOptions Bit Value Identifiers Bit value HexadecimalCommand prompt G fmt opt = g 8b Two place hexadecimal value Fmt Two Place Hexadecimal ValueOpt Two Place Hexadecimal Value Maintenance Cleaning FieldGas Connection Verification Tolerance Chart Gas Temperature 65cAccuracy Drift/week Verification IntervalVerification Procedure Verification complete StepIf the unit fails Verification, you will see the following An average will be calculatedCalibration Procedure CalibrationCalibration Interval Field Calibration Steps Model 2230R, Process Hydrogen Analyzer