Teledyne 2230R instruction manual Fmt opt Low high Low high err not rdy Atm, Format fmt

Page 17

MODEL 2230R, Process Hydrogen Analyzer

E

F

G < fmt> <opt>

H <low> <high>

I <low> <high> <err> <not rdy>

L < fmt> <opt>

P <atm>

S

V <low> <high> <err> <not rdy>

X

6 – Temperature calibration data A – All of the above

Field Verify (refer to the Section 5.4 VERFICATION).

Field Calibration (refer to the Section 5.5 CALIBRATION).

Start or resume the sensor operation: if needed, heat the ASIC, setup the sensor, and output data on serial port; this restores default settings (refer to the FORMAT <fmt> and OPTIONS <opt> sections).

Set the hydrogen reporting range: <low> to <high> in %H.

Set the DAC current output range: <low> to <high>; error output <err>; and not ready output <not rdy> milliamps. Possible range from 0 to 20 milliamps.

Print current hydrogen reading. Used to poll for hydrogen readings. Default format <fmt> is current setting (refer to the FORMAT <fmt> and OPTIONS <opt> sections).

Select atmospheric pressure of gas. Factory default pressure is 1 ATM.

Stop the sensor: turn off heater, set Bias to zero, set DAC outputs to zero, and stop reporting data on the serial output.

Set the DAC voltage output range: <low> to <high>; error output <err>; and not ready output <not rdy> in volts. Possible range from 0 to 5 volts.

Clear field calibration data (returns to last factory calibration data).

Format <fmt>

The Format <fmt> string is a two character hexadecimal representation of an 8 bit value derived from the following table. The user determines which data is needed and selects that bit value. Once all selections are made the values are summed bitwise and then converted to a two place hexadecimal value. To aid in the conversion, a 4 bit to hexadecimal conversion table follows in EXAMPLE 3.

Format: Bit Value Identifiers

Serial output Format <fmt> parameter: select the desired columns of data from this list, add the bit value for each column bitwise, and convert into two hexadecimal characters using the 4 bit-to-Hexadecimal table in EXAMPLE 3.

Description

 

 

 

Bit Value

 

 

 

 

 

 

 

 

 

Include time stamp

1

0

0

0

0

0

0

0

Include raw ADC values

0

1

0

0

0

0

0

0

Include PCB temperature

0

0

1

0

0

0

0

0

Include sensor temperature

0

0

0

1

0

0

0

0

Include capacitor reading

0

0

0

0

1

0

0

0

Page 17

Image 17
Contents 30/2011 ECO # Important NoticesCopyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Teledyne Analytical Instruments DescriptionProcess Hydrogen Analyzer Model 2230R Teledyne InstrumentsDimensions Weight SpecificationsModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Normal operation / Hydrogen Detected below R1 set point Status Indicator ColorHydrogen Detected above R2 set point/ Unit fault detected Wire Color Power/Analog Output ConnectorElectrical & Communication Connections DescriptionAnalog Output Voltage Analog Output Range Power-On Self Diagnostic Error₂ Messages 0.0000 0.0000 Example DisplayExample ESC Format Bit Value Identifiers Fmt opt Low high Low high err not rdy AtmFormat fmt Options opt Calculation Errors Heater StateOptions Bit Value Identifiers Bit value HexadecimalOpt Two Place Hexadecimal Value Command prompt G fmt opt = g 8bTwo place hexadecimal value Fmt Two Place Hexadecimal Value Gas Connection Maintenance CleaningField Verification Tolerance Chart Gas Temperature 65cAccuracy Drift/week Verification IntervalVerification Procedure Verification complete StepIf the unit fails Verification, you will see the following An average will be calculatedCalibration Interval Calibration ProcedureCalibration Field Calibration Steps Model 2230R, Process Hydrogen Analyzer