MODEL 2230R, Process Hydrogen Analyzer
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
| FIELD CALIBRATION STEPS | ||
|
|
|
|
Display |
| User response | |
H2scan: |
|
| Type “ f” to run field calibration |
|
|
|
|
Ready to Calibrate |
|
| Type “ y” |
(Y/N)? |
|
|
|
|
|
|
|
Pressure is 0.9690 |
|
| Type “ y” |
atm (Y/N)? |
|
| Please note that the firmware has remembered the |
|
|
| atmospheric pressure from the preceding Verification |
|
|
| procedure. Type “n” if a change is required and enter the |
|
|
| correct pressure. |
|
|
|
|
Save as default |
|
| Type “ n” |
(Y/N)? |
|
|
|
Gas1 is 0%H2 |
|
| Type “ n” |
(Y/N)? |
|
|
|
Gas1 for res (Y/N)? |
|
| Type “ y” |
Cal Gas: 2.000%H2 |
|
| Type “ n” |
(Y/N)? |
|
|
|
|
|
|
|
Enter gas: |
|
| Type “ 1” for gas #1, 1% hydrogen by volume |
Cal Gas: 1.000%H2 |
|
| Type “ y” |
(Y/N)? |
|
|
|
|
|
|
|
Settle time: 3 min |
|
| Type “ n” |
(Y/N)? n |
|
|
|
Enter time: |
|
| Type “ 30” for a 30 minute duration for gas #1 |
|
|
|
|
Settle time: 30 min |
|
| Type “ y” |
(Y/N)? |
|
|
|
|
|
|
|
Apply 1.000%H2: |
|
| Type “ y” |
Ready (Y/N)? y |
|
|
|
|
|
|
|
14
15
16
17 |
18 |
19 |
20
| Streaming data… |
Taking Average… | Calibration Gas #1 finished |
cap=0.00000 |
|
res=x.xxxxx |
|
Gas2 for res (Y/N)? | Type “ y” |
Cal Gas: | Type “ n” |
100.000%H2 (Y/N)? |
|
|
|
Enter gas: | Type “ 30” for gas #2, 30% hydrogen by volume |
|
|
Cal Gas: 30.000%H2 | Type “ y” |
(Y/N)? |
|
Settle time: 1 min | Type “ n” |
(Y/N)? |
|
|
|
Enter time: | Type “ 30” for a 30 minute duration for gas #2 |
Page 25