Teledyne 2230R instruction manual Field Calibration Steps

Page 25

MODEL 2230R, Process Hydrogen Analyzer

Step

1

2

3

4

5

6

7

8

9

10

11

12

13

 

FIELD CALIBRATION STEPS

 

 

 

 

Display

 

User response

H2scan:

 

 

Type “ f” to run field calibration

 

 

 

 

Ready to Calibrate

 

 

Type “ y”

(Y/N)?

 

 

 

 

 

 

 

Pressure is 0.9690

 

 

Type “ y”

atm (Y/N)?

 

 

Please note that the firmware has remembered the

 

 

 

atmospheric pressure from the preceding Verification

 

 

 

procedure. Type “n” if a change is required and enter the

 

 

 

correct pressure.

 

 

 

 

Save as default

 

 

Type “ n”

(Y/N)?

 

 

 

Gas1 is 0%H2

 

 

Type “ n”

(Y/N)?

 

 

 

Gas1 for res (Y/N)?

 

 

Type “ y”

Cal Gas: 2.000%H2

 

 

Type “ n”

(Y/N)?

 

 

 

 

 

 

 

Enter gas:

 

 

Type “ 1” for gas #1, 1% hydrogen by volume

Cal Gas: 1.000%H2

 

 

Type “ y”

(Y/N)?

 

 

 

 

 

 

 

Settle time: 3 min

 

 

Type “ n”

(Y/N)? n

 

 

 

Enter time:

 

 

Type “ 30” for a 30 minute duration for gas #1

 

 

 

 

Settle time: 30 min

 

 

Type “ y”

(Y/N)?

 

 

 

 

 

 

 

Apply 1.000%H2:

 

 

Type “ y”

Ready (Y/N)? y

 

 

 

 

 

 

 

14

15

16

17

18

19

20

 

Streaming data…

Taking Average…

Calibration Gas #1 finished

cap=0.00000

 

res=x.xxxxx

 

Gas2 for res (Y/N)?

Type “ y”

Cal Gas:

Type “ n”

100.000%H2 (Y/N)?

 

 

 

Enter gas:

Type “ 30” for gas #2, 30% hydrogen by volume

 

 

Cal Gas: 30.000%H2

Type “ y”

(Y/N)?

 

Settle time: 1 min

Type “ n”

(Y/N)?

 

 

 

Enter time:

Type “ 30” for a 30 minute duration for gas #2

Page 25

Image 25
Contents 30/2011 ECO # Important NoticesCopyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Teledyne Analytical Instruments DescriptionProcess Hydrogen Analyzer Model 2230R Teledyne InstrumentsDimensions Weight SpecificationsModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Hydrogen Detected above R2 set point/ Unit fault detected Status Indicator ColorNormal operation / Hydrogen Detected below R1 set point Wire Color Power/Analog Output ConnectorElectrical & Communication Connections DescriptionAnalog Output Voltage Analog Output Range Power-On Self Diagnostic ErrorExample Display₂ Messages 0.0000 0.0000 Example ESC Format fmt Fmt opt Low high Low high err not rdy AtmFormat Bit Value Identifiers Options opt Calculation Errors Heater StateOptions Bit Value Identifiers Bit value HexadecimalTwo place hexadecimal value Fmt Two Place Hexadecimal Value Command prompt G fmt opt = g 8bOpt Two Place Hexadecimal Value Field Maintenance CleaningGas Connection Verification Tolerance Chart Gas Temperature 65cAccuracy Drift/week Verification IntervalVerification Procedure Verification complete StepIf the unit fails Verification, you will see the following An average will be calculatedCalibration Calibration ProcedureCalibration Interval Field Calibration Steps Model 2230R, Process Hydrogen Analyzer