Teledyne 2230R instruction manual Esc

Page 16

MODEL 2230R, Process Hydrogen Analyzer

Serial

Communication

Commands

The unit can be communicated with and configured via the use of commands as described below. Two levels of communication outputs are available:

Level 0 – Default level used for data monitoring and basic functions providing a continuous stream of data readings

Level 1 – Password protected level used for configuration of user-settable parameters; interactive single-line data output per command

Command Summary

The RETURN or ENTER key is the last character of the command string. If either key is pressed without a command string the result is an invalid command and will resume continuous display if in Level 0 or return to prompt if in Level 1.

Keystroke

ESC

sp(spacebar)

A

C

=<password>

Level 0 Commands

Description

Stops continuous display to enter a password or command. If in level zero, the continuous display will resume after executing one command.

Pressing the Space key while the serial output is active will display a label line showing the heading for each column of data.

Average readings.

Clear peak hydrogen value.

Enter the password to change security level. A null or invalid password returns to the default security level.

Level 0 password = “0” Level 1 password = “h2scan

Keystroke

A <R1> <R2>

C

D <page>

Level 1 Commands

Description

Set the R1 and R2 set points for the relay contacts in %H.

Clear peak hydrogen value.

Display Product Information. Enter page number 0-6 or A for all pages, default is page 0.

0 – Product information

1 – User configuration

2 – Manufacturing information

3 – Product configuration

4 – Sensor characterization data

5 – Hydrogen calibration data

Page 16

Image 16
Contents Important Notices 30/2011 ECO #Copyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Description Teledyne Analytical InstrumentsProcess Hydrogen Analyzer Model 2230R Teledyne InstrumentsSpecifications Dimensions WeightModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Hydrogen Detected above R2 set point/ Unit fault detected Status Indicator ColorNormal operation / Hydrogen Detected below R1 set point Power/Analog Output Connector Wire ColorElectrical & Communication Connections DescriptionAnalog Output Power-On Self Diagnostic Error Voltage Analog Output RangeExample Display₂ Messages 0.0000 0.0000 Example ESC Format fmt Fmt opt Low high Low high err not rdy AtmFormat Bit Value Identifiers Calculation Errors Heater State Options optOptions Bit Value Identifiers Bit value HexadecimalTwo place hexadecimal value Fmt Two Place Hexadecimal Value Command prompt G fmt opt = g 8bOpt Two Place Hexadecimal Value Field Maintenance CleaningGas Connection Tolerance Chart Gas Temperature 65c VerificationAccuracy Drift/week Verification IntervalVerification Procedure Step Verification completeIf the unit fails Verification, you will see the following An average will be calculatedCalibration Calibration ProcedureCalibration Interval Field Calibration Steps Model 2230R, Process Hydrogen Analyzer