5. MAINTENANCE

5.1CLEANING

If the unit is exposed to process streams containing debris, condensates or other material that may collect over the sensor tip then the unit should be removed from the stream periodically and the tip cleaned by a gentle wiping with a clean lint-free cloth or paper.

5.2CALIBRATION INTERVAL

TELEDYNE ANALYTICAL INSTRUMENTS recommends that the system calibration be carried out if the unit fails Verification. Field Verification can occur as often as needed and is recommended when the tolerated sensor drift may result in unacceptably high errors. Anticipated drift rates and error calculations are explained in the Verification section of this manual. The user has two choices in calibrating the unit:

 

 

 

Contact TELEDYNE ANALYTICAL INSTRUMENTS to make arrangements

 

Factory

for a comprehensive Factory Calibration using high accuracy gases by

 

TELEDYNE ANALYTICAL INSTRUMENTS’s factory trained technicians;

 

Calibration

 

optional NIST traceable certificate available upon request. An annual

 

 

 

 

 

 

Factory Calibration is advisable to optimize performance.

 

 

 

Refer to Sections 4.4 and 4.5 for the operational procedure. Customer-

 

 

 

specific Field Calibration Kits for the Model 2230R analyzer is available

 

Field

 

from TELEDYNE ANALYTICAL INSTRUMENTS. Field Calibration is a 3 stage

 

 

process:

 

Calibration

 

 

 

1. Run Field Verification,

 

 

 

 

 

 

2. Run Field Calibration,

 

 

 

3. Repeat the Field Verification to confirm a successful Field Calibration.

In the event the Field Calibration procedure seems to not be effective, TELEDYNE ANALYTICAL INSTRUMENTS recommends the unit be shipped back to the factory for a comprehensive evaluation and Factory Calibration.

5.3FIELD VERIFICATION & FIELD CALIBRATION GASES

Field Verification and Field Calibration require the availability of at least two certified known gases at values specific to the user’s application. For optimization, the hydrogen concentrations of the gases implemented in Field Verification and Field Calibration must be the same. Flow rate recommended is 0.5 ± 0.2 slpm.

Verification / Calibration in an Inert (i.e. nitrogen) Background Gas:

For a unit calibrated in an inert gas (i.e. nitrogen) background, carry out Verification and Calibration in an inert (i.e. nitrogen) gas background.

Gas Connection

Gases are applied to the unit through user’s piping.

Page 20

Image 20
Contents Important Notices 30/2011 ECO #Copyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Description Teledyne Analytical InstrumentsProcess Hydrogen Analyzer Model 2230R Teledyne InstrumentsSpecifications Dimensions WeightModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Normal operation / Hydrogen Detected below R1 set point Status Indicator ColorHydrogen Detected above R2 set point/ Unit fault detected Power/Analog Output Connector Wire ColorElectrical & Communication Connections DescriptionAnalog Output Power-On Self Diagnostic Error Voltage Analog Output Range₂ Messages 0.0000 0.0000 Example DisplayExample ESC Format Bit Value Identifiers Fmt opt Low high Low high err not rdy AtmFormat fmt Calculation Errors Heater State Options optOptions Bit Value Identifiers Bit value HexadecimalOpt Two Place Hexadecimal Value Command prompt G fmt opt = g 8bTwo place hexadecimal value Fmt Two Place Hexadecimal Value Gas Connection Maintenance CleaningField Tolerance Chart Gas Temperature 65c VerificationAccuracy Drift/week Verification IntervalVerification Procedure Step Verification completeIf the unit fails Verification, you will see the following An average will be calculatedCalibration Interval Calibration ProcedureCalibration Field Calibration Steps Model 2230R, Process Hydrogen Analyzer