MODEL 2230R, Process Hydrogen Analyzer
Now use the 4
two place hexadecimal value: |
|
<fmt> Two Place Hexadecimal Value : | 8b |
From the OPTIONS <opt> table above, you identify your desired columns with its corresponding bit value:
<opt> Descriptions | <opt> Bit Value |
Calculations Errors | 0000 0100 |
Heater State | 0000 0010 |
<opt> 4 Bit Value Combination: | 0000 0110 |
Again, use the 4
<opt> Two Place Hexadecimal Value : | 06 |
Conclusion: To have the Time Stamp, Capacitor Reading, Overall Hydrogen Reading, the Peak Hydrogen Reading, Calculations Errors and Heater State columns continuously reported, you will implement the “G” serial command (“Go” Command, refer to LEVEL 1 COMMANDS table) as follows at the H2scan:
command prompt: G < fmt> <opt> = g 8b 06
Display | User Response |
H2scan: | Type “g 8b 06” |
Page 19