Teledyne 2230R instruction manual Two place hexadecimal value Fmt Two Place Hexadecimal Value

Page 19

MODEL 2230R, Process Hydrogen Analyzer

Now use the 4 BIT-TO-HEXADECIMALtable above to convert this 4 bit value combination into a

two place hexadecimal value:

 

<fmt> Two Place Hexadecimal Value :

8b

From the OPTIONS <opt> table above, you identify your desired columns with its corresponding bit value:

<opt> Descriptions

<opt> Bit Value

Calculations Errors

0000 0100

Heater State

0000 0010

<opt> 4 Bit Value Combination:

0000 0110

Again, use the 4 BIT-TO-HEXADECIMALtable to convert this 4 bit value combination into a two place hexadecimal value:

<opt> Two Place Hexadecimal Value :

06

Conclusion: To have the Time Stamp, Capacitor Reading, Overall Hydrogen Reading, the Peak Hydrogen Reading, Calculations Errors and Heater State columns continuously reported, you will implement the “G” serial command (“Go” Command, refer to LEVEL 1 COMMANDS table) as follows at the H2scan:

command prompt: G < fmt> <opt> = g 8b 06

Display

User Response

H2scan:

Type “g 8b 06”

Page 19

Image 19
Contents 30/2011 ECO # Important NoticesCopyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Teledyne Instruments DescriptionTeledyne Analytical Instruments Process Hydrogen Analyzer Model 2230RDimensions Weight SpecificationsModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Hydrogen Detected above R2 set point/ Unit fault detected Status Indicator ColorNormal operation / Hydrogen Detected below R1 set point Description Power/Analog Output ConnectorWire Color Electrical & Communication ConnectionsAnalog Output Voltage Analog Output Range Power-On Self Diagnostic ErrorExample Display₂ Messages 0.0000 0.0000 Example ESC Format fmt Fmt opt Low high Low high err not rdy AtmFormat Bit Value Identifiers Bit value Hexadecimal Calculation Errors Heater StateOptions opt Options Bit Value IdentifiersTwo place hexadecimal value Fmt Two Place Hexadecimal Value Command prompt G fmt opt = g 8bOpt Two Place Hexadecimal Value Field Maintenance CleaningGas Connection Verification Interval Tolerance Chart Gas Temperature 65cVerification Accuracy Drift/weekVerification Procedure An average will be calculated StepVerification complete If the unit fails Verification, you will see the followingCalibration Calibration ProcedureCalibration Interval Field Calibration Steps Model 2230R, Process Hydrogen Analyzer