MODEL 2230R, Process Hydrogen Analyzer
Step
1
2
3
4
5
6
7
8 |
9 |
10 |
11 |
12
13
14
15
16
17
| FIELD VERIFICATION STEPS | ||
|
|
|
|
Display |
| User response | |
H2scan: |
|
| Type “ e” to run field verification |
|
|
|
|
Ready to Verify |
|
| Type “ y” |
(Y/N)? |
|
|
|
|
|
|
|
Pressure is 1.000 |
|
| Type “ n” |
atm (Y/N)? |
|
|
|
|
|
|
|
Enter new pressure: |
|
| Type “ .969” for the local atmospheric pressure |
Pressure is 0.9690 |
|
| Type “ y” |
atm (Y/N)? |
|
|
|
|
|
|
|
Save as default |
|
| Type “ n” |
(Y/N)? |
|
|
|
|
|
|
|
Verify Sensor |
|
| Type “ y” |
(Y/N)? |
|
|
|
|
|
|
|
Enter Gas (%H2): |
|
| Type “ 1” for gas #1, 1% hydrogen by volume |
Settle time (min): |
|
| Type “ 30” for a 30 minute duration for gas #1 |
|
|
|
|
Tolerance (%H2): |
|
| Type “0.3%” absolute from the for 1% hydrogen by volume |
|
|
|
|
Apply 1.000%H2: |
|
| Type “ y” |
Ready (Y/N)? |
|
|
|
|
|
| Streaming data… |
Taking Average… |
|
| An average will be calculated. Verification Gas1 finished |
|
|
|
|
Continue test Verify |
|
| Type “ y” |
Sensor (Y/N)? |
|
|
|
|
|
|
|
Enter Gas (%H2): |
|
| Type “ 30” for gas #2, 30% hydrogen by volume |
Settle time (min): |
|
| Type “ 30” for a 30 minute duration for gas #2 |
Tolerance (%H2): |
|
| Type “1.0%” absolute from the TOLERANCE CHART for 30% |
|
|
| hydrogen by volume |
|
|
|
|
Apply 30.000%H2: |
|
| Type “ y” |
Ready (Y/N)? |
|
|
|
|
|
|
| Streaming data… |
|
|
|
|
|
|
|
18 | Taking Average… | An average will be calculated. Verification Gas2 finished. |
|
19Continue test Verify Type “n” to end field verification.
|
| Sensor (Y/N)? | Verification complete. |
4). | If the unit passes Verification, Calibration is not required at this time. | ||
5). | If the unit fails Verification, you will see the following: | ||
|
|
|
|
18 |
| Taking Average… | An average will be calculated. |
19 |
| Verify Failed | Verification complete. |
6). | If the unit fails Verification, TELEDYNE ANALYTICAL INSTRUMENTS recommends | ||
| Calibration, detailed in the following section. |