Teledyne 2230R instruction manual Step, Verification complete, An average will be calculated

Page 23

MODEL 2230R, Process Hydrogen Analyzer

Step

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

 

FIELD VERIFICATION STEPS

 

 

 

 

Display

 

User response

H2scan:

 

 

Type “ e” to run field verification

 

 

 

 

Ready to Verify

 

 

Type “ y

(Y/N)?

 

 

 

 

 

 

 

Pressure is 1.000

 

 

Type “ n

atm (Y/N)?

 

 

 

 

 

 

 

Enter new pressure:

 

 

Type “ .969” for the local atmospheric pressure

Pressure is 0.9690

 

 

Type “ y

atm (Y/N)?

 

 

 

 

 

 

 

Save as default

 

 

Type “ n

(Y/N)?

 

 

 

 

 

 

 

Verify Sensor

 

 

Type “ y

(Y/N)?

 

 

 

 

 

 

 

Enter Gas (%H2):

 

 

Type “ 1” for gas #1, 1% hydrogen by volume

Settle time (min):

 

 

Type “ 30” for a 30 minute duration for gas #1

 

 

 

 

Tolerance (%H2):

 

 

Type “0.3%” absolute from the for 1% hydrogen by volume

 

 

 

 

Apply 1.000%H2:

 

 

Type “ y

Ready (Y/N)?

 

 

 

 

 

 

Streaming data…

Taking Average…

 

 

An average will be calculated. Verification Gas1 finished

 

 

 

 

Continue test Verify

 

 

Type “ y

Sensor (Y/N)?

 

 

 

 

 

 

 

Enter Gas (%H2):

 

 

Type “ 30” for gas #2, 30% hydrogen by volume

Settle time (min):

 

 

Type “ 30” for a 30 minute duration for gas #2

Tolerance (%H2):

 

 

Type “1.0%” absolute from the TOLERANCE CHART for 30%

 

 

 

hydrogen by volume

 

 

 

 

Apply 30.000%H2:

 

 

Type “ y

Ready (Y/N)?

 

 

 

 

 

 

 

Streaming data…

 

 

 

 

 

 

 

18

Taking Average…

An average will be calculated. Verification Gas2 finished.

 

19Continue test Verify Type “n” to end field verification.

 

 

Sensor (Y/N)?

Verification complete.

4).

If the unit passes Verification, Calibration is not required at this time.

5).

If the unit fails Verification, you will see the following:

 

 

 

 

18

 

Taking Average…

An average will be calculated.

19

 

Verify Failed

Verification complete.

6).

If the unit fails Verification, TELEDYNE ANALYTICAL INSTRUMENTS recommends

 

Calibration, detailed in the following section.

Page 23

Image 23
Contents 30/2011 ECO # Important NoticesCopyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Teledyne Instruments DescriptionTeledyne Analytical Instruments Process Hydrogen Analyzer Model 2230RDimensions Weight SpecificationsModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Normal operation / Hydrogen Detected below R1 set point Status Indicator ColorHydrogen Detected above R2 set point/ Unit fault detected Description Power/Analog Output ConnectorWire Color Electrical & Communication ConnectionsAnalog Output Voltage Analog Output Range Power-On Self Diagnostic Error₂ Messages 0.0000 0.0000 Example DisplayExample ESC Format Bit Value Identifiers Fmt opt Low high Low high err not rdy AtmFormat fmt Bit value Hexadecimal Calculation Errors Heater StateOptions opt Options Bit Value IdentifiersOpt Two Place Hexadecimal Value Command prompt G fmt opt = g 8bTwo place hexadecimal value Fmt Two Place Hexadecimal Value Gas Connection Maintenance CleaningField Verification Interval Tolerance Chart Gas Temperature 65cVerification Accuracy Drift/weekVerification Procedure An average will be calculated StepVerification complete If the unit fails Verification, you will see the followingCalibration Interval Calibration ProcedureCalibration Field Calibration Steps Model 2230R, Process Hydrogen Analyzer