MODEL 2230R, Process Hydrogen Analyzer

21

22

23

Settle time: 30 min

Type “ y”

(Y/N)?

 

 

 

Apply 30.000%H2:

Type “ y”

Ready (Y/N)?

 

 

Streaming data…

Taking Average…

Calibration Gas #2 finished.

cap=0.00000

Calibration complete.

res=x.xxxxx

 

4)Per TELEDYNE ANALYTICAL INSTRUMENTS’s recommendation, upon completion of Calibration, conduct the Verification sequence as described in the previous section to verify that the calibration process was executed correctly.

Page 26

Image 26
Contents Important Notices 30/2011 ECO #Copyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Process Hydrogen Analyzer Model 2230R DescriptionTeledyne Analytical Instruments Teledyne InstrumentsSpecifications Dimensions WeightModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Normal operation / Hydrogen Detected below R1 set point Status Indicator ColorHydrogen Detected above R2 set point/ Unit fault detected Electrical & Communication Connections Power/Analog Output ConnectorWire Color DescriptionAnalog Output Power-On Self Diagnostic Error Voltage Analog Output Range₂ Messages 0.0000 0.0000 Example DisplayExample ESC Format Bit Value Identifiers Fmt opt Low high Low high err not rdy AtmFormat fmt Options Bit Value Identifiers Calculation Errors Heater StateOptions opt Bit value HexadecimalOpt Two Place Hexadecimal Value Command prompt G fmt opt = g 8bTwo place hexadecimal value Fmt Two Place Hexadecimal Value Gas Connection Maintenance CleaningField Accuracy Drift/week Tolerance Chart Gas Temperature 65cVerification Verification IntervalVerification Procedure If the unit fails Verification, you will see the following StepVerification complete An average will be calculatedCalibration Interval Calibration ProcedureCalibration Field Calibration Steps Model 2230R, Process Hydrogen Analyzer