Teledyne 2230R instruction manual Calibration Interval, Calibration Procedure

Page 24

MODEL 2230R, Process Hydrogen Analyzer

5.5CALIBRATION

Calibration Interval

Calibration is used to correct any offset that exist between the sensor output and a known hydrogen concentration. Calibrations do not cause any wear on the sensor and can be accomplished as often as desired. It is recommended that Calibration be performed if a unit fails Verification.

Calibration Procedure

IMPORTANT NOTE: During the Calibration process any previously completed Field Calibrations are cancelled. As a result, during the routine the unit may display a hydrogen concentration that is different from the applied gas concentration. This is normal. Once the procedure is completed, the readings will be corrected to display the right concentrations for all subsequent exposures.

Calibration can only be accomplished through interface with the unit via the serial port (refer to the previous Serial Communication Commands section and the LEVEL 1 COMMANDS table in Section 4.3). Analog outputs can be monitored through the user’s system.

Calibration Using Serial Interface (firmware version 0.47) – Follow the sequence below:

1)Press “Esc

2)The unit will return the command prompt “H2scan:

3)Type “f” then hit “Enter” and follow prompts to field calibrate the sensor with two gas

concentrations

These steps will implement the following specification values as an example; for values in bold, please substitute appropriate values relative to your specific operational conditions:

Local atmospheric pressure: 0.969 ATM

Hydrogen range: 0% to 30% hydrogen, balance nitrogen

Field Verification Gas #1: 1% hydrogen, balance nitrogen (instead of gas #1 being 0% hydrogen, H2scan recommends Field Verification gases always have a hydrogen concentration)

Field Verification Gas #2: 30% hydrogen, balance nitrogen

Settle Time: 30 minutes (TELEDYNE ANALYTICAL INSTRUMENTS recommends this duration for field calibration)

WARNING: AS IN THIS EXAMPLE, FOR OPTIMIZATION, THE HYDROGEN CONCENTRATIONS OF THE GASES IMPLEMENTED IN FIELD CALIBRATION AND FIELD VERIFICATION MUST BE THE SAME.

Page 24

Image 24
Contents Important Notices 30/2011 ECO #Copyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Description Teledyne Analytical InstrumentsProcess Hydrogen Analyzer Model 2230R Teledyne InstrumentsSpecifications Dimensions WeightModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Status Indicator Color Hydrogen Detected above R2 set point/ Unit fault detectedNormal operation / Hydrogen Detected below R1 set point Power/Analog Output Connector Wire ColorElectrical & Communication Connections DescriptionAnalog Output Power-On Self Diagnostic Error Voltage Analog Output RangeDisplay Example₂ Messages 0.0000 0.0000 Example ESC Fmt opt Low high Low high err not rdy Atm Format fmtFormat Bit Value Identifiers Calculation Errors Heater State Options optOptions Bit Value Identifiers Bit value HexadecimalCommand prompt G fmt opt = g 8b Two place hexadecimal value Fmt Two Place Hexadecimal ValueOpt Two Place Hexadecimal Value Maintenance Cleaning FieldGas Connection Tolerance Chart Gas Temperature 65c VerificationAccuracy Drift/week Verification IntervalVerification Procedure Step Verification completeIf the unit fails Verification, you will see the following An average will be calculatedCalibration Procedure CalibrationCalibration Interval Field Calibration Steps Model 2230R, Process Hydrogen Analyzer