Teledyne 2230R Electrical & Communication Connections, Power/Analog Output Connector, Wire Color

Page 12

MODEL 2230R, Process Hydrogen Analyzer

If the unit gets exposed to extended periods of no H(or some Oxygen), condition the sensor as described in the Start-up section and follow-up with a Field Verification (and Field Calibration, if needed).

Effect of pressure: The Model 2230R Analyzer is hydrogen specific and sensitive to only the hydrogen partial pressure in the gas stream. Since changes in total gas pressure will affect the hydrogen partial pressure, they will also affect the sensor readings. For instance, at one atmosphere pressure, a 50% H/Nmixture will be reported as 50% from the unit. At 1.1 atm, the reading will increase to 55% and two atms will result in a reading of 100%. In fact, the Model 2230R Analyzer are capable of measuring multiple atmospheres of hydrogen and readings above 100% Hare interpreted as hydrogen pressures above one atmosphere. So, for example, a reading of 150% Hmeans 1.5 times the hydrogen pressure of a 100% Hconcentration at one atmosphere. At the factory, the units are calibrated at one atmosphere pressure. Performing the Field Calibration at the operating pressure will display the pressure corrected reading. For example, if the local atmospheric pressure is 0.97 atm, doing a Field Calibration will correct for this.

Verification/Calibration gas bottle accuracy: The inaccuracy of the gas bottle concentration will directly affect the measured accuracy by the units. During factory calibration, the units are calibrated with high accuracy gases (as high as ± 0.02% accurate). It is strongly recommended that the user perform calibration with similar high accuracy gases to maintain the accuracy specified in the manual.

4. ELECTRICAL & COMMUNICATION

WARNING: IF THE UNIT IS INSTALLED IN A CLASSIFIED LOCATION THEN IT IS THE RESPONSIBILITY OF THE USER AND INSTALLER TO MAKE CONNECTIONS TO RELATED EQUIPMENT IN A MANNER CONSISTENT WITH THE LOCATION CLASSIFICATION.

4.1CONNECTIONS

Power/Analog Output - Connector 1

Supplied Cable – 4m (12 ft.) standard length (Other lengths available)

Wire Color

Brown

White

Black

Blue

Description

+8 VDC to +13 VDC VDC Return

Positive Analog Output Analog Output Return

Page 12

Image 12
Contents Important Notices 30/2011 ECO #Copyright 2006 Teledyne Analytical Instruments Model 2230R, Process Hydrogen Analyzer Safety Messages Table of Contents Description Teledyne Analytical InstrumentsProcess Hydrogen Analyzer Model 2230R Teledyne InstrumentsSpecifications Dimensions WeightModel 2230R, Process Hydrogen Analyzer Operation Unit Location Process Connection Status Indicator Color Hydrogen Detected above R2 set point/ Unit fault detectedNormal operation / Hydrogen Detected below R1 set point Power/Analog Output Connector Wire ColorElectrical & Communication Connections DescriptionAnalog Output Power-On Self Diagnostic Error Voltage Analog Output RangeDisplay Example₂ Messages 0.0000 0.0000 Example ESC Fmt opt Low high Low high err not rdy Atm Format fmtFormat Bit Value Identifiers Calculation Errors Heater State Options optOptions Bit Value Identifiers Bit value HexadecimalCommand prompt G fmt opt = g 8b Two place hexadecimal value Fmt Two Place Hexadecimal ValueOpt Two Place Hexadecimal Value Maintenance Cleaning FieldGas Connection Tolerance Chart Gas Temperature 65c VerificationAccuracy Drift/week Verification IntervalVerification Procedure Step Verification completeIf the unit fails Verification, you will see the following An average will be calculatedCalibration Procedure CalibrationCalibration Interval Field Calibration Steps Model 2230R, Process Hydrogen Analyzer