3M eg4, eg3 user manual Dosimeter fields explained for the Edge

Models: eg4 eg3

1 74
Download 74 pages 7.74 Kb
Page 31
Image 31

25

Customizing set-up parameters in DMS

Dosimeter setting

Dosimeter settings for the Edge

Name Field

 

 

*Note: refer to Glossary for more detailed information.

 

Heartbeat enable

When enabled (or checked), the LED will flash green while running

 

 

and flash amber when stopped.

 

Response

Either Slow Time Response or Fast Time Response is applied to

 

 

the measurements. (Typically Slow time response is used in the

 

 

US market while Fast is used in the EU communities.)

 

C-A enable

When enabled (or checked), C minus A will be calculated.

 

Projected Time

A selectable time parameter via the up/down arrows or type in a

 

 

value. The projected time value is used to make projected dose

 

 

calculation for the dose measurement.

 

Peak Weighting

Select either A, C or Z for the peak frequency weighting.

 

 

 

 

Dose indicator LED

When enabled (or checked), the LED will flash red when the dose

 

 

value exceeds the designated entered value.

 

Setup name

(Optional field) A setup name may be entered into the Meter 1

 

 

and/or Meter 2 fields. When set, it appears on the top of Edge’s

 

 

display (example setup names: OHSHA, MSHA, EU directive).

 

Threshold enable

When the checkbox is selected and a value is entered for meter 1

 

 

and/or meter 2, the threshold is used to calculate Lavg, TWA, and

 

 

 

 

Dose measurements.

 

Upper Limit

Select a UL dB level by the up/down arrows or by typing in a value.

 

 

(Note: The upper limit (UL) is a measurement setting in which the

 

 

user selects a specific decibel level. The UL value records the

 

 

amount of time that the Sound level is at or greater than the

 

 

entered level.)

 

RMS Weighting

Select either A or C RMS Weighting. (Note: The RMS (root mean

 

 

square) are the results from the Edge with the response time and

 

 

weighting settings taken into account.)

 

Criterion Level

Select a Criterion level via the up/down arrow or by typing in a

 

 

value. (Note: Criterion level is the average SPL that will result in

 

 

100% dose over the criterion time and is typically set by a

 

 

regulatory agency such as OSHA, EU directive.)

 

Exchange Rate

Enter in (or select with up/down arrows) 3, 4, or 5 dB. ER refers to

 

 

how sound energy is averaged over time. (OSHA uses a 5 dB

 

 

doubling rate while most of the world uses 3 dB.)

 

Table 1-2: Dosimeter fields explained for the Edge

053-644, Rev E

Page 25

eg4 & eg3

Page 31
Image 31
3M eg4, eg3 user manual Dosimeter fields explained for the Edge