Operation

Operation

Introduction

Introduction

This section provides information on how to make effective use of the 11970 mixers.

Operating Precautions

Refer to the sections below for specific parameters to follow prior to mixer operation.

WARNING Do not exceed the maximum ratings listed below or permanent damage to the mixer will result.

RF Input Power

Use the following parameters:

CW: No greater than 20 dBm

Pulse: No greater than 24 dBm at < 1 mSec

Average: No greater than 20 dBm

LO Input Power

Make sure the LO input power is no greater than 20 dBm.

Electrostatic Discharge

When installing the mixer, you must always connect the SMA cables to the spectrum analyzer and LO amplifier BEFORE connecting them to the mixer. This will minimize the danger of an electrostatic discharge damaging the mixer diodes.

11975A ALC Switch

BEFORE using the 11975A Amplifier to increase the LO input power, set the amplifier ALC switch to the ON position. When this switch is in the OFF position the LO power can be greater than 20 dBm. This level of LO power can destroy the mixer diodes. The ALC switch is on the amplifier rear panel.

Waveguide Protective Foam

Do not remove, displace, or damage the white, nonconductive foam installed in the open end of the waveguide. Since the mixer is amplitude calibrated with this foam in place, tampering with it affects the calibration.

18

Chapter 2

Page 18
Image 18
Agilent Technologies 11970-90031 manual Introduction, Operating Precautions