Performance Tests

Performance Tests

Performance Test Procedures

Table 3-4. Recommended Test Equipment for 11970U

Instrument

Critical Specifications

Recommended Model

 

 

 

Spectrum Analyzer

LO and IF ranges compatible with mixer

8566B

Synthesized Sweeper

Frequency: 13.3 to 20 GHz

8340A

 

Output Level: > 9 dBm

 

Amplifier

Output Level: > 18 dBm leveled

11975A

 

Frequency Range: 4 to 6 GHz

 

Amplifier

Frequency Range: 13.3 to 20 GHz

8349A

 

Output Level >15 dBm

 

Power Meter

Compatible with Power Sensor

432A

Thermistor Mount

SWR: <2.0

Hughes1 45773H-1100

Thermistor Mount

SWR: <1.3 @ 6 GHz

478A

Directional Coupler2

Coupling: 20 dB

U752D

 

Directivity: >30 dB

 

 

Auxiliary Arm SWR: <1.2

 

Isolator

Insertion Loss: <2 dB

U365A

 

Isolation: >20 dB

 

 

SWR: <1.5

 

Frequency Tripler

Conversion Loss: <20 dB

Spacekom Microwave3

 

 

TKa-1

Adapter

SMA female to Type N female

P/N 1250-1404

Adapter

SMA female to SMA female

P/N 1250-1158

Adapter (2 required)

Type N male to SMA female

P/N 1250-1250

Cables (4 required)

Connectors: SMA male

P/N 5061-5458

Cable

Connectors: SMA male

P/N 8120-4396

 

Loss: <1.0 dB @ 20 GHz

 

Cable

Connectors: BNC male

11086A

 

 

 

1.Hughes Aircraft Co. Electron Dynamics Division, Torrance, CA

2.Calibration data for the coupling ration between the output arm and the auxil- iary arm is necessary for accurate measurements.

3.Honeywell, Inc. Spacekom Microwave Center, Santa Barbara, CA