Agilent Technologies 11970-90031 Amplitude Calibration, Manually Entering Conversion-Loss Data

Models: 11970-90031

1 78
Download 78 pages 56.4 Kb
Page 23
Image 23

Operation

Using the Mixers with the E4407B Spectrum Analyzer (Option AYZ)

2.On the analyzer, press Preset, Factory Preset, if present.

3.Select external mixing by pressing Input/Output, Input Mixer, Input Mixer (Ext).

4.The analyzer frequency band will be set to 26.5 - 40 GHz (A). To choose a different band, press Ext Mix Band , then press the desired band frequency range/letter key. For this example, we will use band A, which ranges from 36.5 GHz to 40 GHz.

Amplitude Calibration

The conversion loss versus frequency data can be entered on your analyzer from one of three sources.

From a Conversion Loss Data Disk, supplied with your mixer. See “Using a Conversion-Loss Data Disk with the ESA or PSA Series Analyzers” on page 21 .

Data from the calibration sheet supplied with your mixer.

From conversion-loss data located on the mixer body label.

Manually Entering Conversion-Loss Data

1.The analyzer frequency band will be set to 26.5 – 40 GHz (A). To choose a different band, press Ext Mix Band and then press the desired band frequency range/letter key. For this example, we will use band A, which ranges from 26.5 GHz to 40 GHz.

NOTE

To correct for the conversion-loss of the harmonic mixer in use, the analyzer

 

amplitude correction feature is used.

 

 

2.Press AMPLITUDE Y Scale, More , Corrections. Select a correction set for use with external mixing. The recommended set to use is Other although any available set could be used.

3.Press Edit to enter the appropriate conversion loss data for the mixer in use. These values are listed on the mixer, or a calibration sheet that is supplied with the mixer.

NOTE

More correction points entered across the band in use will improve frequency

 

response accuracy. Up to 200 points may be defined for each set.

 

 

4.Once the desired correction points are entered, press Return, Correction (On) to activate correction set Other. This will also turn corrections on resulting in a calibrated display. It is recommended that the correction set entered be saved on the internal memory or the floppy drive for future reference. See the ESA User’s and Programmer’s Reference guide for information on saving correction values.

Operation

Chapter 2

23

Page 23
Image 23
Agilent Technologies 11970-90031 manual Amplitude Calibration, Manually Entering Conversion-Loss Data