Operation

Using the Mixers with the 856X Series Spectrum Analyzers

In this example, we’ll look at U-band, which ranges from 40 GHz to 60 GHz, as shown in Figure 2-9. The LOCK HARMONIC function “locks” the spectrum analyzer in that band, ensuring that the spectrum analyzer sweeps only the chosen band. LOCK ON OFF is automatically set to ON when FULL BAND is pressed.

Figure 2-9 Select the band of interest.

Operation

Save the average conversion-loss value

4.Default conversion-loss values that are stored in the analyzer for each frequency band are listed in Table 2-1. These values approximate the values for the Agilent 11970 series mixers. Other conversion-loss values may be entered into the spectrum analyzer in two ways. The first method lets you to save the average conversion-loss value for the entire band using AVERAGE CNV LOSS. To activate this function:

a.Press AUX CTRL, EXTERNAL MIXER, AMPTD CORRECT, AVERAGE CNV LOSS.

b.Enter the appropriate average conversion-loss value. On Agilent 11970 Series harmonic mixers, these values are printed on the mixer. The U-band mixer used for this example had an average conversion loss of 23.5 dB, as shown in Figure 2-10.

32

Chapter 2

Page 32
Image 32
Agilent Technologies 11970-90031 manual Save the average conversion-loss value, Select the band of interest