Operation

Using the Mixers with the E4440A, E4446A, or E4448A PSA Series Spectrum Analyzer (Option AYZ)

Using the Mixers with the E4440A, E4446A, or E4448A PSA Series Spectrum Analyzer (Option AYZ)

Be sure to refer to “Spectrum Analyzer Retrofit Requirements” on page 9 for instrument requirements for the PSA Series spectrum analyzers. The Agilent Technologies PSA Series spectrum analyzers contain an extensive menu of functions that help with millimeter measurements. The following examples explain how to connect external mixers to the spectrum analyzer, how to choose the band of interest, how to store and activate conversion-loss factors, and how to use the signal-identification functions.

Set up the equipment

1. Connect the signal source and harmonic mixer to the analyzer, as shown in Figure 2-2.

Figure 2-5 Harmonic Mixer Setup

Spectrum Analyzer

IF INPUT

SMA Cable

1st LO OUTPUT

SMA Cable

IF

LO

Operation

 

RF Input

SIGNAL

Agilent 11970 SERIES

SOURCE

HARMONIC MIXER

unpremix5

NOTE

Agilent 5061-5458 SMA type cables should be used to connect the mixer IF

 

and LO ports to the analyzer. Do not over-tighten the cables. The maximum

 

torque should not exceed 112 N-cm (10 in-lb.)

 

 

26

Chapter 2

Page 26
Image 26