Agilent Technologies 87130A manual CYCLes?, Syntax, Description, Example Command, Remote Operation

Models: 87130A

1 140
Download 140 pages 20.39 Kb
Page 75
Image 75
:CYCLes?

Remote Operation

:CYCLes?

:CYCLes?

Syntax

DIAGnostics:EERom:CYCLes?

Description

This query causes the 87130A to read back the number of times to which the EEROM has been written. If this number exceeds 10,000 the EEROM should be replaced. If the EEROM has never been written to, then 0 is returned.

Example Command

DIAG:EEROM:CYCLES?

Agilent 87130A Operating and Service Manual 4-33

Page 75
Image 75
Agilent Technologies 87130A manual Syntax, Description, Example Command, Remote Operation, DIAGnosticsEERomCYCLes?