Remote Operation

:DELete

:DELete

Syntax

:PATH

<path name>

ROUTE

:DELete <group name>

:GROUP

 

 

[:ALL]

Description

This command deletes:

All data associated with the specified path or group and frees up the path storage register.

All channels from the PFAIL list.

This command sets memory to an initial state.

NOTE

The :DELete and the :DELay shortform is the same. The :DELete command

 

 

applies only in syntax with other subsystems and not directly to channel lists

 

 

from ROUTe.

 

 

 

Example Command

ROUTE:PATH:DELETE ATTEN_14;

 

 

This deletes the path with pathname ATTEN_14 and all data in it.

 

 

ROUTE:PATH:DELETE:ALL;

This deletes all paths.

ROUTE:GROUP:DEL ATTEN;

This deletes all data associated with the group ATTEN. The group name is set to the default (refer to MEM:DELETE).

ROUTE:GROUP:DEL:ALL;

This deletes all data associated with all the groups. The group names are set to the default (refer to MEM:DELETE).

ROUTe:PFAil:DELete;

This command removes all data from the PFAIL:OPEN and PFAIL:CLOSE lists.

4-38Agilent 87130A Operating and Service Manual

Page 80
Image 80
Agilent Technologies 87130A manual DELete

87130A specifications

Agilent Technologies 87130A is a highly regarded network analyzer that has been widely adopted in the field of RF and microwave testing. Known for its versatility and precision, the 87130A is designed for evaluating components and devices in various applications, including telecommunications and electronics.

One of the main features of the 87130A is its frequency range, which spans from 300 kHz to 3 GHz. This broad spectrum allows engineers to test a wide variety of RF components, including amplifiers, filters, and antennas. The analyzer provides vector network analysis capabilities, enabling users to measure both the magnitude and phase of S-parameters, essential for characterizing components effectively.

The instrument is equipped with a powerful internal calibration capability, which ensures accurate measurements by compensating for test fixture errors and other variables. The built-in electronic calibration options further enhance measurement accuracy and repeatability, giving engineers confidence in their results.

Another notable characteristic of the 87130A is its graphical user interface, which is intuitive and user-friendly. The display provides clear visual representations of measurement data, allowing for easy interpretation and analysis. Users can access a variety of measurement functions, such as gain compression, group delay, and return loss, making it a versatile tool for multi-faceted testing.

The Agilent 87130A employs advanced measurement algorithms that improve speed and accuracy. These enhancements facilitate rapid data acquisition, which is particularly beneficial in production environments where time is critical. The analyzer also supports multiple calibration techniques, including through-reflect-match (TRM) and short-open-load (SOL), catering to diverse testing needs.

Connectivity options are abundant, with GPIB, LAN, and USB interfaces, allowing seamless integration into automated test setups. This modularity supports remote operation, making it easier to incorporate the analyzer into various testing environments.

Overall, Agilent Technologies 87130A stands out for its precision, ease of use, and flexibility in RF and microwave testing applications. Its combination of advanced features and robust performance makes it a staple in laboratories, production lines, and research settings, helping engineers and technicians achieve optimal results in their measurement endeavors.