Agilent Technologies 87130A manual Execution Errors, Device Specific Errors

Models: 87130A

1 140
Download 140 pages 20.39 Kb
Page 86
Image 86
Execution Errors

Remote Operation

:ERRor?

-181,Invalid outside macro definition

-183,Invalid inside macro definition

Execution Errors

These set the Exec Err bit in ESR.

-200,Execution error -222,Data out of range -223,Too much data -270,Macro error -272,Macro execution error -273,Illegal macro label -276,Macro recursion error

-277,Macro redefinition not allowed

Device Specific Errors

These set the Cmd Err bit in ESR.

-310,System error

-310,System error;SCPI internal -310,System error;no memory -310,System error;formatter 777,Queue Overflow

1001,Sense error;<17 hex digits>

This error is generated when an apparently impossible state of the sense lines from one or more relays is detected. The first of the 17 hex digits represents the card number on which the failure was detected; the rest represent a 64-bit binary number which indicates which of the 32-channels experienced failures.

Each channel has two associated bits; the leftmost indicates that the error was detected on the CLOSed side of the relay; the rightmost that the error was detected on the OPEN side. A failure is indicated with a 1. The least significant (rightmost) two bits are channel 0, the most significant (leftmost) are channel 31. If more than one card experiences failures then more than one error will be reported.

1002,Memory capacity exceeded

An attempt has been made to store more data than will fit in the 16 Kbyte nonvolatile memory area.

1003,Timer unstable

Instability has been detected in the 6840 timer chip, so the operating system cannot run reliably.

4-44Agilent 87130A Operating and Service Manual

Page 86
Image 86
Agilent Technologies manual Execution Errors, Device Specific Errors, Agilent 87130A Operating and Service Manual