TRIGger

Remote Operation

TRIGger

TRIGger

Syntax

TRIGger:SEQuence:DELay <number>

Description

This command allows the user to control the 24 V power supply recovery time. Input values are limited to between 0 and 200 msec (200 msec is the default setting). The GPIB command TRIGger:SEQuence:DELay will allow entry of a new power recovery time. Values outside the 0 to 200 msec range will return an error.

Example Command

TRIG:SEQ:DEL .15

This example assigns the value, 150 msec, to the TRIGger command.

Example Query

TRIG:SEQ:DEL?

 

The query form returns the assigned number. For the example, it would

 

return

 

.15

 

 

NOTE

The default 200msec TRIG:SEQ:DEL value will be reset on instrument power

 

up or if the *RST command is issued.

 

 

4-62Agilent 87130A Operating and Service Manual