Agilent Technologies 87130A manual STATus, Syntax, Description

Models: 87130A

1 140
Download 140 pages 20.39 Kb
Page 102
Image 102
STATus

Remote Operation

STATus

STATus

Syntax

:[EVENt]?

:CONDition

STATus:OPERation :ENABle[?]

:PTRansition

:NTRansition[?]

:[EVENt]?

STATus:QUEStionable :CONDition?

:ENABle[?]

Description

The STATus subsystem is fully specified by SCPI. Refer to the following

commands: *STB (Status Byte), *SRE (Request Enable), *ESE (Event Status Enable), and *ESR? (Event Status Register Query)

STATus:OPERation

In the STATus:OPERation subsystem, only the settling bit (bit 1) is implemented by the 87130A; it is set during a switching operation (and when saving to EEROM) and cleared when it completes. The “SWITCHING” LED is on whenever this bit is set. The following commands are applicable:

STATus:OPERation:EVENt

STATus:OPERation:CONDITION?

STATus:OPERation:ENABle

STATus:OPERation:PTRansition[?]

STATus:OPERation:NTRansition[?]

STATus:OPERation:PTRansition[?]

STATus:QUEStionable

The QUEStionable susbsytem is required by SCPI but is not used by the 87130A, 70612A,C or 70613A,C. The following commands are implemented:

STATus:QUEStionable:EVENt?

STATus:QUEStionable:CONDition?

STATus:QUEStionable:ENABle

4-60Agilent 87130A Operating and Service Manual