GPIB Addressing

GPIB Addressing

Installing the 87130A Attenuator/Switch Driver

Getting Started

3.Use the five-bit binary address switches located on the rear panel of the instrument to change the GPIB address.

The 87130A has a factory preset address of 9. There are 32 possible addresses. The switch labeled with a one is the least significant bit.

Addresses 0 and 31 are typically reserved for GPIB functions and should not be used. IEEE-488.1 limits the number of addressable elements (instruments) to 16.

Figure 2-1 87130A GPIB Address Switch (Default 9 shown)

Agilent 87130A Operating and Service Manual 2-3