Chapter 8 Specifications

Performance Specifications

Performance Specifications

Output Ratings(@0°C - 40°C)

Low range

0 to +15 V/0 to 7 A

High range

0 to +30 V/0 to 4 A

Programming Accuracy[1] 12 months(@25°C ± 5°C), ±(% of output + offset)

Voltage

0.05% + 10 mV

Current

0.2% + 10 mA

Readback Accuracy[1] 12 months(over GPIB and RS-232 or front panel with respect to actual output @25°C ± 5°C), ±(% of output + offset)

Voltage

0.05% + 5 mV

Current

0.15% + 5 mA

Ripple and Noise (with outputs ungrounded, or with either output terminal grounded, 20 Hz to 20 MHz)

Normal mode voltage

<0.35 mV rms and 2 mV p-p

Normal mode current

<2 mA rms

Common mode current

<1.5 mA rms

Load Regulation, ±(% of output + offset)

Change in output voltage or current for any load change within ratings with remote sensing connected

Voltage

<0.01%

+ 2 mV

Current

<0.01%

+ 250 mA

Line Regulation, ±(% of output + offset)

Change in output voltage and current for any line change within ratings

Voltage

<0.01%

+ 2 mV

Current

<0.01%

+ 250 mA

[1]Accuracy specifications are after an 1-hour warm-up with no load and calibration at 25 °C.

157

8

Page 158
Image 158
Agilent Technologies E3632A manual Specifications Performance Specifications, 157