Volt/Curr

Display Limit

Output On/Off

Chapter 3 Front-Panel Operation

Constant Current Operation

5Adjust the knob for the desired output current. 1

Check that the Limit annunciator still blinks. Set the knob for current control. The second digit of the ammeter will be blinking. Change the blinking digit using the resolution selection keys and adjust the knob to the desired output current.

6Return to the meter mode.

Press Display Limit key or let the display time-out after several seconds to return the meter mode. Notice that the Limit annunciator turns off and the display shows “OUTPUT OFF” message.

7Enable the output.

3

The OFF annunciator turns off and the 15V or 30V, OVP, OCP and CC annunciators are lit. Notice that the display is in the meter mode. In the meter mode, the display shows the actual output voltage and current.

Refer to “Programming Overvoltage Protection” and “Programming Overcurrent Protection” sections, starting on page 42 and page 45 for more information on OVP and OCP annunciators.

8Verify that the power supply is in the constant current mode.

If you operate the power supply in the constant current (CC) mode, verify that the CC annunciator is lit. If the CV annunciator is lit, choose a higher voltage limit.

N o t e

During actual CC operation, if a load change causes the voltage limit to be

 

exceeded, the power supply will automatically crossover to constant voltage

 

mode at the preset voltage limit and the output current will drop

 

proportionately.

 

 

 

 

Remote interface operation:

 

 

VOLTage {<voltage>MINMAX}

Set the voltage

 

CURRent {<current>MINMAX}

Set the current

 

OUTPut ON

Enable the output

1You can use the resolution selection keys to move the blinking digit to the right or left when setting the current.

39

Page 40
Image 40
Agilent Technologies E3632A manual Verify that the power supply is in the constant current mode