I/O Config

Chapter 3 Front-Panel Operation

Remote Interface Configuration

To Set the Baud Rate and Parity (RS-232)

To configure the power supply for the RS-232 interface, proceed as follows:

1Turn on the remote configuration mode.

I/O Config

GPIB / 488

You will see the above message on the display if the power supply has not been changed from the factory setting.

Notice that if you changed the remote interface selection to RS-232 before, “RS-232” message will be displayed.

2Choose the RS-232 interface.

RS-232

You can choose the RS-232 interface by turning the knob to the left.

3Move to the RS-232 interface setting mode and select the baud rate.

3

I/O Config

9600 BAUD

The rate is set to 9600 baud when the power supply is shipped from the factory. Choose from one of the following by turning the knob to the right or left: 300, 600, 1200, 2400, 4800, or 9600 baud.

4Save the change and choose the parity.

NONE 8 BITS

The power supply is configured for 8 data bits with no parity when shipped from the factory. Choose from one of the following by turning the knob to the right or left: None 8 Bits, Odd 7 Bits, or Even 7 Bits. When you set parity, you are indirectly setting the number of the data bits.

59

Page 60
Image 60