Chapter 4 Remote Interface Reference

System-Related Commands

*TST?

This query performs a complete self-test of the power supply. Returns ‘‘0’’ if the self-test passes or ‘‘1’’ or any non-zero value if it fails. If the self-test fails, an error message is also generated with additional information on why the test failed.

*SAV { 1 2 3 }

This command stores the present state of the power supply to the specified location in non-volatile memory. Three memory locations (numbered 1, 2 and

3)are available to store operating states of the power supply. The state storage feature ‘‘remembers’’ the states or values of the following commands:

 

CURR, CURR:STEP, CURR:TRIG, CURR:PROT, CURR:PROT:STAT DISP,

 

OUTP, OUTP:REL, TRIG:DEL, TRIG:SOUR, VOLT, VOLT:STEP, VOLT:TRIG,

 

VOLT:PROT, VOLT:PROT:STAT, and VOLT:RANG

 

To recall a stored state, you must use the same memory location used

 

previously to store the state.

 

*RCL { 1 2 3 }

 

This command recalls a previously stored state. To recall a stored state, you

 

must use the same memory location used previously to store the state.

 

 

N o t e

DISP {OFFON} can be stored and recalled in remote interface mode only.

 

Going to local mode automatically sets the display state to ON.

 

 

95

4

Page 96
Image 96
Agilent Technologies E3632A manual SAV 1 2, RCL 1 2