Chapter 3 Front-Panel Operation

Remote Interface Configuration

Remote Interface Configuration

Before you can operate the power supply over the remote interface, you must configure the power supply for the remote interface. This section gives information on configuring the remote interface. For additional information on programming the power supply over the remote interface, See "Remote Interface Reference", starting on page 71 in chapter 4.

Remote Interface Selection

The power supply is shipped with both an GPIB (IEEE-488) interface and an RS-232 interface on the rear panel. Only one interface can be enabled at a time. The GPIB interface is selected when the power supply is shipped from the factory.

The remote interface can be selected from the front-panel only.

The interface selection is stored in non-volatile memory, and does not change when power has been off or after a remote interface reset.

If you select the GPIB interface, you must select a unique address for the

power supply. The current address is displayed momentarily on the front panel when you turn on the power supply.1

Your GPIB bus controller has its own address. Be sure to avoid using the bus controller’s address for any instrument on the interface bus. Agilent Technologies controllers generally use address “21”.

If you enable the RS-232 interface, you must select the baud rate and parity

to be used. “RS-232” is displayed momentarily on the front panel when you turn on the power supply if you have selected this interface.2

1Refer to "GPIB Interface Configuration" starting on page 61 for more information on connecting the power supply to a computer over the GPIB interface.

2Refer to "RS-232 Interface Configuration" starting on page 62 for more information on connecting the power supply to a computer over the RS-232 interface.