Agilent Technologies E4400-90326 Creating a Step Sweep and a List Sweep, Configuring a Step Sweep

Models: E4400-90326

1 31
Download 31 pages 360 b
Page 7
Image 7

ESG Family Signal Generators

Getting Started

 

Creating a Step Sweep and a List Sweep

Creating a Step Sweep and a List Sweep

This section will show you two ways to set up the signal generator to sweep a defined set of points. You will create a ten-point step sweep with the settings listed below, then use these points as the basis for a new list sweep.

frequency range from 525 MHz to 600 MHz

power level from −20 dBm to 0 dBm

dwell time 500 ms at each point

Configuring a Step Sweep

1.

Press Preset.

This sets the signal generator to its factory-defined instrument

 

 

state.

2.

Press Sweep/List.

The first level of sweep softkeys is displayed.

3.

Press the Sweep Type List Step

This toggles the Sweep Type List Step softkey from List to

 

softkey.

Step.

4.

Press the Configure Step Sweep

Another menu is displayed with softkeys that you will use to

 

softkey.

create the sweep points.

5.Press the Freq Start softkey. This sets the new start frequency for the step sweep to 525 MHz.

Enter 525 using the numeric keypad and press the MHz softkey.

6.Press the Freq Stop softkey. Enter 600 using the numeric keypad and press the MHz softkey.

This sets the new stop frequency for the step sweep to 600 MHz.

7.Press the Ampl Start softkey. Enter 20 using the numeric keypad and press the dBm softkey.

This sets the power level for the start of the step sweep to 20 dBm.

Quick Start Guide

1-5

Page 7
Image 7
Agilent Technologies E4400-90326 quick start Creating a Step Sweep and a List Sweep, Configuring a Step Sweep, State