Getting Started

ESG Family Signal Generators

Creating a Step Sweep and a List Sweep

 

8.Press the Ampl Stop softkey. Enter 0 using the numeric keypad and press the dBm softkey.

9.Press the # Points softkey. Enter 10 by rotating the front panel knob until the number 10 is displayed.

10.Press the Step Dwell softkey. Enter 500 using the numeric keypad and pressing the msec softkey.

This sets the new power level for the end of the step sweep to 0 dBm.

This sets the number of points in the step sweep.

This sets the dwell time for each point in the step sweep.

Turning On Continuous Step Sweep

1. Press Return.

This moves the softkey menu up one level.

2.Press the Sweep softkey.Another menu is displayed, showing you choices for sweeping either the frequency, amplitude, or frequency and amplitude data.

3.Press the Freq & Ampl softkey. Selecting this softkey returns you to the previous menu and turns the sweep function on.

4.

Press the

This toggles the Sweep Repeat Single Cont softkey from

 

Sweep Repeat Single Cont

Single to Cont (Continuous). Notice that the SWEEP display

 

softkey.

annunciator is turned on, indicating that the signal generator is

 

 

sweeping.

5.

Press RF On/Off.

Notice that the display annunciator changes from RF OFF to

 

 

RF ON. The swept RF signal is now available at the

 

 

RF OUTPUT connector.