Appendix B Verification and Calibration

Test Record Form – Agilent N5744A and N5764A

Agilent N5744A and N5764A

Report No _______________

Date __________________

Description

Model

Minimum Specs.

Results

Maximum Specs.

Voltage Programming & Readback

 

 

 

 

 

Minimum Voltage Vout

Both

0 mV

__________

+ 50 mV

Measurement Readback

Both

Vout 20 mV

__________

Vout + 20 mV

High Voltage Vout

Both

19.98 V

__________

20.02 V

Measurement Readback

Both

Vout 40 mV

__________

Vout + 40 mV

CV Load Effect

Both

4 mV

__________

+ 4 mV

CV Source Effect

Both

4 mV

__________

+ 4 mV

CV Ripple and Noise

 

 

 

 

 

peak-to-peak

Both

N/A

__________

60 mV

rms

Both

N/A

__________

8 mV

Transient Response

 

 

 

 

 

Voltage @ 1ms

Both

100 mV

__________

+ 100 mV

 

 

 

 

 

 

Current Programming & Readback

 

 

 

 

 

Minimum Current Iout

N5744A

0 mA

__________

+ 191 mA

 

N5764A

0 mA

__________

+ 381 mA

Measurement Readback

N5744A

Iout 114 mA

__________

Iout + 114 mA

 

N5764A

Iout 228 mA

__________

Iout + 228 mA

High Current Iout

N5744A

37.924 A

__________

38.076 A

 

N5764A

75.848 A

__________

76.152 A

Measurement Readback

N5744A

Iout 152 mA

__________

Iout + 152 mA

 

N5764A

Iout 304 mA

__________

Iout + 304 mA

CC Load Effect

N5744A

12.6 mA

__________

+ 12.6 mA

 

N5764A

20.2 mA

__________

+ 20.2 mA

CC Source Effect

N5744A

5.8 mA

__________

+ 5.8 mA

 

N5764A

9.6 mA

__________

+ 9.6 mA

 

 

 

 

 

 

 

 

Test Description

 

N5744A Setting

N5764A Setting

Voltage Programming & Readback, Min Voltage

40 mV, 152 mA

40 mV, 304 mA

Voltage Programming & Readback, High Voltage

20 V, 152 mA

 

 

20 V, 304 mA

CV Load Effect, Source Effect, Ripple and Noise

20 V, 38 A

 

 

20 V, 76 A

Transient Response

 

20 V, from 3.8 A to 34.2 A

20 V, from 7.6 A to 68.4 A

Current Programming & Readback, Min Current

152 mA, 20 V

 

 

304 mA, 20 V

Current Programming & Readback, High Current

38 A, 20 V

 

 

76 A, 20 V

CC Load Effect, Source Effect

 

38 A, 20 V

 

 

76 A, 20 V

 

 

 

 

 

 

 

 

 

 

 

 

Load Requirements

 

N5744A

 

 

N5764A

Current shunt

 

0.01100 A

 

 

0.01100 A

Agilent N3300 Electronic load modules

 

2 – N3305A

 

 

3 – N3305A

Fixed Resistor for CV Ripple and Noise

 

0.531kW

 

 

0.262kW

104

Series N5700 User’s Guide

Page 104
Image 104
Agilent Technologies N5700 manual Test Record Form Agilent N5744A and N5764A

N5700 specifications

Agilent Technologies, now part of Keysight Technologies, is renowned for its innovative solutions in electronic measurement and instrumentation. Among its impressive portfolio, the Agilent N5700 series of power supplies stands out, providing precision, reliability, and versatility for a range of applications in test and measurement.

The N5700 series features multiple models, each designed to meet the varying power requirements of test systems and electronic devices. One of its key characteristics is its adjustable output voltage and current, allowing users to set parameters according to specific test needs. With outputs ranging from 0 to 60 V and up to 6 A, the N5700 series caters to both low and high-power applications effectively.

A notable technology integrated into the N5700 series is its advanced measurement capabilities. The built-in voltmeter and ammeter enable real-time monitoring of output voltage and current, ensuring precise control over the power supplied to the device under test. This feature is particularly advantageous in troubleshooting and optimization scenarios, providing engineers with immediate feedback on performance.

The N5700 also embraces the latest in power supply control technology with its intuitive user interface. The front panel incorporates a clear display and simple navigation controls, allowing users to program settings easily and access functions without extensive training. Additionally, remote programming capabilities via GPIB, USB, or LAN facilitate integration into automated test setups, enhancing productivity and efficiency.

Thermal management is another hallmark of the N5700 series. Its design ensures effective heat dissipation, enabling reliable operation even under demanding conditions. This robustness is critical in environments where consistent performance is essential, such as in research laboratories and manufacturing facilities.

Furthermore, safety features are thoughtfully included in the N5700 series, such as overvoltage protection, overcurrent protection, and short-circuit protection. These elements assure users that their devices and test setups remain secure during testing, minimizing the risk of damage.

In summary, the Agilent N5700 series power supplies encapsulate the fusion of precision, advanced measurement technology, user-friendly design, thermal management, and robust safety features. This combination makes them an invaluable tool for engineers and researchers engaged in electronic testing and development across various industries. With its versatility and reliability, the N5700 series continues to play a pivotal role in advancing electronic measurement methodologies.