Agilent Technologies N5700 Rack Installation, Cleaning, Than 6 mm into the sides of the unit

Models: N5700

1 132
Download 132 pages 28.95 Kb
Page 18
Image 18

2 Installation

Rack Installation

 

Ensure that the screws used to attach the rack slide kit do not penetrate more

CAUTION

 

than 6 mm into the sides of the unit.

 

Do not block the air intake at the front, or the exhaust at the rear of the unit.

The Agilent N5700 power supplies can be mounted in a standard 19- inch rack panel or cabinet. They are designed to fit in one rack unit

(1U) of space. To install the power supply in a rack:

1.Use the front panel rack-mount brackets to install the power supply in the rack.

2.Use a support bracket to provide adequate support for the rear of the power supply.

3.If using rack mount slides, use Agilent N5740A Rack-mount Slide Kit to install the unit in a standard 19-inch equipment rack. Refer to the following figure for assembly instructions. Use two #10-32 x 3/8 in (max.) screws at each side. To prevent internal damage, use the specified screw length only.

Cleaning

WARNING SHOCK HAZARD To prevent electric shock, unplug the unit before cleaning.

Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.

18

Series N5700 User’s Guide

Page 18
Image 18
Agilent Technologies N5700 manual Rack Installation, Cleaning, Than 6 mm into the sides of the unit