Verification and Calibration Appendix B

Test Record Form – Agilent N5749A and N5769A

Agilent N5749A and N5769A

Report No _______________

Date __________________

Description

Model

Minimum Specs.

Results

Maximum Specs.

Voltage Programming & Readback

 

 

 

 

 

Minimum Voltage Vout

Both

0 mV

__________

+ 250 mV

Measurement Readback

Both

Vout 100 mV

__________

Vout + 100 mV

High Voltage Vout

Both

99.9 V

__________

100.1 V

Measurement Readback

Both

Vout 200 mV

__________

Vout + 200 mV

CV Load Effect

Both

12 mV

__________

+ 12 mV

CV Source Effect

Both

12 mV

__________

+ 12 mV

CV Ripple and Noise

 

 

 

 

 

peak-to-peak

Both

N/A

__________

80 mV

rms

Both

N/A

__________

8 mV

Transient Response

 

 

 

 

 

Voltage @ 1ms

Both

500 mV

__________

+ 500 mV

 

 

 

 

 

 

Current Programming & Readback

 

 

 

 

 

Minimum Current Iout

N5749A

0 mA

__________

+ 38 mA

 

N5769A

0 mA

__________

+ 76 mA

Measurement Readback

N5749A

Iout 22.5 mA

__________

Iout + 22.5 mA

 

N5769A

Iout 45 mA

__________

Iout + 45 mA

High Current Iout

N5749A

7.485 A

__________

7.515 A

 

N5769A

14.97 A

__________

15.03 A

Measurement Readback

N5749A

Iout 30 mA

__________

Iout + 30 mA

 

N5769A

Iout 60 mA

__________

Iout + 60 mA

CC Load Effect

N5749A

6.5 mA

__________

+ 6.5 mA

 

N5769A

8 mA

__________

+ 8 mA

CC Source Effect

N5749A

2.75 mA

__________

+ 2.75 mA

 

N5769A

3.5 mA

__________

+ 3.5 mA

 

 

 

 

 

 

 

 

Test Description

 

N5749A Setting

N5769A Setting

Voltage Programming & Readback, Min Voltage

200 mV, 30 mA

200 mV, 60 mA

Voltage Programming & Readback, High Voltage

100 V, 30 mA

 

 

100 V, 60 mA

CV Load Effect, Source Effect, Ripple and Noise

100 V, 7.5 A

 

 

100 V, 15 A

Transient Response

 

100 V, from 0.75 A to 6.75 A

100 V, from 1.5 A to 13.5 A

Current Programming & Readback, Min Current

30 mA, 100 V

 

 

60 mA, 100 V

Current Programming & Readback, High Current

7.5 A, 100 V

 

 

15 A, 100 V

CC Load Effect, Source Effect

 

7.5 A, 100 V

 

 

15 A, 100 V

 

 

 

 

 

 

 

 

 

 

 

 

Load Requirements

 

N5749A

 

 

N5769A

Current shunt

 

0.115 A

 

 

0.115 A

Agilent N3300 Electronic load modules

 

2 – N3305A

 

 

3 – N3305A

Fixed Resistor for CV Ripple and Noise

 

13.331kW

 

 

6.672kW

Series N5700 User’s Guide

109

Page 109
Image 109
Agilent Technologies N5700 manual Test Record Form Agilent N5749A and N5769A

N5700 specifications

Agilent Technologies, now part of Keysight Technologies, is renowned for its innovative solutions in electronic measurement and instrumentation. Among its impressive portfolio, the Agilent N5700 series of power supplies stands out, providing precision, reliability, and versatility for a range of applications in test and measurement.

The N5700 series features multiple models, each designed to meet the varying power requirements of test systems and electronic devices. One of its key characteristics is its adjustable output voltage and current, allowing users to set parameters according to specific test needs. With outputs ranging from 0 to 60 V and up to 6 A, the N5700 series caters to both low and high-power applications effectively.

A notable technology integrated into the N5700 series is its advanced measurement capabilities. The built-in voltmeter and ammeter enable real-time monitoring of output voltage and current, ensuring precise control over the power supplied to the device under test. This feature is particularly advantageous in troubleshooting and optimization scenarios, providing engineers with immediate feedback on performance.

The N5700 also embraces the latest in power supply control technology with its intuitive user interface. The front panel incorporates a clear display and simple navigation controls, allowing users to program settings easily and access functions without extensive training. Additionally, remote programming capabilities via GPIB, USB, or LAN facilitate integration into automated test setups, enhancing productivity and efficiency.

Thermal management is another hallmark of the N5700 series. Its design ensures effective heat dissipation, enabling reliable operation even under demanding conditions. This robustness is critical in environments where consistent performance is essential, such as in research laboratories and manufacturing facilities.

Furthermore, safety features are thoughtfully included in the N5700 series, such as overvoltage protection, overcurrent protection, and short-circuit protection. These elements assure users that their devices and test setups remain secure during testing, minimizing the risk of damage.

In summary, the Agilent N5700 series power supplies encapsulate the fusion of precision, advanced measurement technology, user-friendly design, thermal management, and robust safety features. This combination makes them an invaluable tool for engineers and researchers engaged in electronic testing and development across various industries. With its versatility and reliability, the N5700 series continues to play a pivotal role in advancing electronic measurement methodologies.