5 Language Reference
Output Commands
Output commands enable the output,
OUTPut[:STATe] ONOFF
OUTPut[:STATe]?
This command enables or disables the specified output(s). The enabled state is On (1); the disabled state is Off (0). The state of a disabled output is a condition of zero output voltage and a zero source current (see *RST). The query returns 0 if the output is off, and 1 if the output is on. The *RST value = Off.
OUTPut:PON:STATe RSTAUTO
OUTPut:PON:STATe?
This command determines if the
Refer to *RST and *RCL under System Commands for details.
OUTPut:PROTection:CLEar
This command clears the latched signals that have disabled the output. The
All conditions that generate the fault must be removed before the latch can be cleared. The output is then restored to the state it was in before the fault condition occurred.
68 | Series N5700 User’s Guide |