Using the Script Menu MT8850A / MT8852A Bluetooth Test Set
9. Press the key to switch between DH1, DH3 and DH5.
10. Press the
[1 of 3] soft key to move to the second screen.
11. Choose “Packets per step”. Use the keypad to enter a value for the number of
packers that are averaged at each power step.
12. Choose “Minimum Test power”. Use the keypad to enter a value for the minimum
EUT power. When a power below this value is received by the MT8850A/MT8852A it
will start sending increment power commands, even if it has not received a minimum
power command from the EUT.
13. Choose “Measurement delay”. Use the keypad to specify the time delay between the
step command and the time that measurement is commenced.
14. Press the
[2 of 3] soft key to move to the third screen.
15. The third screen will vary according to whether the Test type is in Loopback or TX
mode. The following screen is in Loopback mode.
The following screen is in TX mode.
You can select or deselect the test to run and make changes to the frequency values
shown.
16. To select or deselect a test, choose the test and press the key.
17. To change a frequency, choose the frequency, press the key, enter the value
using the keypad and press the [Enter] soft key.
18. Press the
[Defaults] soft key to restore the default values.
13000-000109 8-15
June 2003 Testing in Standard Script Mode