MT8850A / MT8852A Bluetooth Test Set Using the Script Menu
12. Press the [1 of 2] soft key to move to the second screen.
The second screen will vary according to whether the Test type is in Loopback or TX
mode. The following screen is in Loopback mode.
The following screen is in TX mode.
You can select or deselect the test to run and make changes to the frequency values
shown.
13. To select or deselect a test, choose the test and press the key.
14. To change a frequency, choose the frequency, press the key, enter the value
using the keypad and press the [Enter] soft key.
15. Press the
[Defaults] soft key to restore the default values.
16. Press the
[Limits] soft key to display the Carrier drift test limits screen.
17. Press the
[Defaults] soft key to restore the default values.
18. Choose the limit you want to change.
19. Press the key and enter the new limit using the keypad.
20. Press the
[Enter] soft key.
21. Press the
[Conditions] soft key to return to the test conditions screen.
22. Press the key to return to the Script menu.
8-22 13000-000109
Testing in Standard Script Mode June 2003