APPENDIX “D” - PROGRAMMABLE FUNCTIONS
Programming of the indicator is divided into modular steps. Each module is a short sequence of data entries. The front panel buttons “UP” and “DOWN” (shown as “arrows” on the front panel) are used to change the data and
“Pro 0” - RETURN TO MEASUREMENT MODE
“Pro 1” - PROGRAM THERMISTOR TYPE, TEMPERATURE SCALE
AND DECIMAL POINT POSITON
“tYPE” | - Select thermistor type (400/700) | |
“SCALE” | - | Enter either degree (F or C) |
“dECPNt” | - | Enter resolution (0 or 0.0) |
“Pro 2” - PROGRAM TEMPERATURE SLOPE AND OFFSET
“SLOPE” - Enter display slope (0.0001 to 9.9999) “OFFSEt” - Enter offset
“Pro 3” - PROGRAM FUNCTIONS ACCESSIBLE WITH FRONT PANEL LOCKOUT ENGAGED
“dSP AL” | - | Enable display alarms |
“ENt AL” | - Enable enter alarms | |
“dSPHYS” | - | Enable display hysteresis |
“ENtHYS” | - Enable enter hysteresis | |
“rSt AL” | - Enable reset latched alarms | |
“dSPbUF” | - Enable display of peak/valley readings | |
“rStbUF” | - Enable reset of peak/valley readings | |
“SELdSP” | - Enable switching display between input and total | |
“rSttOt” | - | Enable reset total |
“dSPOFF” | - Enable display offset value | |
“ENtOFF” | - Enable enter offset value |
“Pro 4” - PROGRAM DIGITAL FILTERING AND REMOTE INPUT FUNCTION
“FILter” | - Enter level of digital filtering |
0- no digital filtering
1- normal filtering
2- increased filtering
3- maximum filtering
0- offset input to zero
1- reset total
2- reset and gate totalizer
3- gate totalizer
4- display hold
5- reset peak/valley
6- reset peak and start peak indicator
7- reset valley and start valley indicator
8- reset latched alarms
9- reset all alarms
10- toggle display between input and total
11- offset input to zero and totalizer the offset values
12- display hold with tare
13- instrument reading with synchronization
14- print request
This sequence may be subject to being