mar345 Program Manual | 1 5 |
______________________________________________________________________________________
Table continued from previous page.
_______________________________________________________________________
ItemDescription
________________________________________________________________________
First image no. | Image number of first image. Range is 1 through 999. |
No. of images | Number of images to be collected. Range is 1 through 999. |
Exp. time | Exposure time in seconds. In DOSE mode, the actual time |
| may be different from the desired exposure time. |
Oscillations | Number of ϕ− oscillations per image. One oscillation is |
| stronlgy recommended for exposure times < 15 min. |
∆ ϕ per image | ϕ movement per image. |
Starting ϕ | ϕ position to start data collection. |
Distance | Distance crystal to detector. |
________________________________________________________________________
For convenience, the program calculates the maximum resolution possible from the chosen distance and the chosen scan mode (image plate diameter).
The image names are created according to the following convention:
directory / root_TYPE_XXX.extension
−TYPE is:
not used in Single Data Set mode
1, 2, 3 or 4 in Multiple Data Set mode
INDEX | in Index Crystal mode |
A or B | in MAD Data mode |
e.g. /images/zeo_001.mar1200 e.g. /images/zeo_1_001.mar1200 e.g. /images/zeo_INDEX_001.mar1200 e.g. /images/zeo_A_001.mar1200
−XXX is a three digit image number like 027,
−extension is:mar1200, mar1600, mar1800, mar2000, mar2300, mar2400, mar3000 or mar345 for Output format "mar345", depending on the Scan mode chosen.
In Index Crystal mode "First image number" is replaced by "ϕ increment" (see above). The starting image will always be 001 and the following images will have numbers that will be calculated from their ϕ −position relative to image number 001 according to:
XXX = i * ϕ increment / ∆ ϕ + 1 | (i=0,1,2,3...) |
e.g. with ϕ increment = 90.0 and ∆ ϕ = 2.0, XXX = 1, 46, 91, | 136, etc. |