advanced menu controls
EN
| Display Example |
| Mode | Description | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Firmware Version |
|
| All | Shows version of installed firmware. |
| V0.01 01/06/2006 |
|
| ||
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| Serial Number |
|
| All | Shows the serial number of the device. |
| HD123456 |
|
| ||
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| SmartCodeTM is used for therapy management |
| SmartCode |
|
|
| |
|
|
|
| reporting; therapy information is encoded in a series | |
| FF28D9 |
|
| All | |
|
|
| of letters and numbers, and includes a special | ||
|
|
|
|
| character for error checking to ensure accuracy. |
|
|
|
|
|
|
Intellipap Enable Menu List
The Enable menu is available in all modes and is a
| Display Example | Description | ||
|
|
|
|
|
|
|
| This option controls the | |
|
| |||
|
| starts the flow of air after breathing once or twice into the | ||
| Enabled |
| ||
|
| mask. | ||
|
|
| ||
|
|
|
|
|
|
|
|
| This option controls the |
|
|
| automatically stops the flow of air after taking off the mask. | |
| Enabled |
| The device may require up to 30 seconds to turn OFF | |
|
|
| depending on the kind of mask used. | |
|
|
|
|
|
|
|
| This option controls the mask fit check feature. The air flow to | |
|
|
| ||
|
|
|
| your mask is constantly measured and if the amount of airflow |
| Mask Fit Check |
|
| |
|
|
| exceeds 95 liters/minute for more than 10% of the time used, | |
| Enabled |
|
| |
|
| a | notification will be created. This notification will be displayed | |
|
|
| the next time you use the IntelliPAP. |
Exit Menu Press
Pressing the up key exits the Enable menu.